Aims: Although several studies have shown positive outcomes after the use of drug-coated balloons (DCB) for in-stent restenosis (ISR), data on randomised controlled trials versus latest-generation drug-eluting stents (DES) are limited. Therefore, in this randomised trial, we sought to evaluate whether a butyryl-tri-hexyl citrate (BTHC)-based paclitaxel DCB is non-inferior to a biodegradable polymer sirolimus-eluting stent (BP-SES) therapy in patients with ISR in either a bare metal stent (BMS) or DES.
Methods And Results: A total of 229 patients with ISR in BMS or DES from 13 German centres and one Latvian centre were 2:1 randomly allocated to DCB (n=157) or DES (n=72). The primary efficacy endpoint was defined as in-stent late lumen loss (LLL) at six months, and the primary safety endpoint was target lesion failure (TLF) at 12 months. LLL in the DCB arm was 0.03±0.40 mm compared to 0.20±0.70 mm in the DES arm (p=0.40). DCB proved to be non-inferior to DES (Δ = -0.17±0.52 mm, 97.5% CI -∞; -0.01]; p<0.0001). At 12 months, Kaplan-Meier TLF estimates were 16.7% in the DCB arm and 14.2% in the DES arm (p=0.65) and remained similar at 18 months (DCB versus DES: 17.4% versus 19.5%, p=0.88).
Conclusions: In patients with DES or BMS ISR, treatment with a paclitaxel DCB showed similar LLL at six months and TLF rates up to 18 months compared to a second-generation sirolimus DES.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4244/EIJ-D-17-01079 | DOI Listing |
Catheter Cardiovasc Interv
January 2025
State Key Laboratory of Frigid Zone Cardiovascular Diseases, Cardiovascular Research Institute and Department of Cardiology, General Hospital of Northern Theater Command, Shenyang, China.
Background: Biodegradable polymer stents may reduce the risk of neoatherosclerosis and stent thrombosis. Limited data is available for biodegradable polymer sirolimus-eluting stent (BP-SES) and durable polymer drug-eluting stents (DP-EES) in chronic total occlusions (CTO).
Aim: This study was to evaluate healing patterns of BP-SES versus DP-EES in CTO at 3 and 13 months based on optical coherence tomography (OCT).
Am Heart J
January 2025
Department of Cardiology, Odense University Hospital, Odense, Denmark; University of Southern Denmark, Odense, Denmark.
Rationale: The biodegradable polymer BioMatrix Alpha™ stent contains biolimus A9 drug which is sirolimus derivative increase in lipophicity. The biodegradable polymer sirolimus eluting Combo™ stent is a dual-therapy sirolimus-eluting and CD34+ antobody coated stent capturing endothelial progenitor cells (EPCs).
Hypothesis: The main hypothesis of the SORT OUT XI trial was that the biodegradable polymer biolimus A9 BioMatrix Alpha ™ stent is noninferior to the biodegradable polymer sirolimus eluting Combo™ stent in an all-comers population with coronary artery disease undergoing percutaneous coronary intervention (PCI).
Am J Cardiol
January 2025
Department of Cardiovascular Medicine, Osaka University Graduate School of Medicine, Suita, Japan.
A dual-therapy sirolimus-eluting and CD34+ antibody-coated Combo Stent (DTS) has been developed to enhance endothelization and capture endothelial progenitor cells; however, vessel responses following DTS implantation remain unclear. Therefore, we evaluated early- and mid-term intravascular characteristics of DTS using intravascular imaging modalities. This multicenter, prospective, observational study enrolled 88 patients (95 lesions) who underwent DTS (43 patients, 48 lesions) or sirolimus-eluting Orsiro stent (SES, 45 patients, 47 lesions) implantation.
View Article and Find Full Text PDFFront Med (Lausanne)
January 2025
Department of Nephrology and Dialysis, Brugmann University Hospital, Université Libre de Bruxelles (ULB), Brussels, Belgium.
Background: The arteriovenous fistula (AVF) is the preferred vascular access for hemodialysis. AVF stenosis is a common complication, often requiring balloon angioplasty. For recurrent stenosis, AVF stenting may be an option.
View Article and Find Full Text PDFJACC Cardiovasc Interv
January 2025
Department of Cardiology, Fu Wai Hospital, National Center for Cardiovascular Diseases of China, Beijing, China. Electronic address:
Background: First-generation bioresorbable scaffolds (BRS) increased risks of stent thrombosis and adverse events. The Bioheart scaffold is a new poly-L-lactic acid-based BRS.
Objectives: This study sought to evaluate the efficacy and safety of the BRS in patients with coronary artery disease.