Background: In myocardial perfusion imaging (MPI), single-photon emission tomography (SPECT) soft-tissue attenuation by the abdomen, breasts, and lateral chest wall may create artifacts that mimic true perfusion defects. This may cause misdiagnosis of myocardial perfusion. The aim of the present study was to compare the localization, extent, and depth of attenuation artifacts in MPI SPECT for a multi-pinhole cadmium zinc telluride (CZT) camera vs a conventional gamma camera.

Methods: Phantom and patient measurements were performed using a CZT camera (GE NM 530c) and a conventional gamma camera (GE Ventri). All images were attenuation corrected with externally acquired low-dose computed tomography. The localization, extent, and depth of the attenuation artifact were quantified by comparing attenuation-corrected and non-attenuation-corrected images.

Results: Attenuation artifacts were shifted from the inferolateral wall to the lateral wall using the CZT camera compared to a conventional camera in both the patient and the phantom. The extent of the attenuation artifact was significantly larger for the CZT camera compared to the conventional camera (23 ± 5% vs 15 ± 5%, P < .001) for patients and the result was similar for the phantom (28% vs 19%). Furthermore, the depth of the attenuation artifact (percent of maximum counts) was less pronounced for the CZT camera than for the conventional camera, both for phantom measurements (73% vs 67%) and patients (72 ± 3% vs 68 ± 4%, P < .001).

Conclusions: Attenuation artifacts are found in different locations to different extents and depths when using a CZT camera vs a conventional gamma camera for MPI SPECT. This should be taken into consideration when evaluating MPI SPECT studies to avoid misinterpretation of myocardial perfusion distribution.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6908561PMC
http://dx.doi.org/10.1007/s12350-018-1296-6DOI Listing

Publication Analysis

Top Keywords

czt camera
16
conventional gamma
12
myocardial perfusion
8
localization extent
8
extent depth
8
depth attenuation
8
attenuation artifacts
8
attenuation artifact
8
camera compared
8
compared conventional
8

Similar Publications

Diagnostic Accuracy of Low-Dose Myocardial Perfusion Imaging in a Real-World Setting.

J Nucl Cardiol

January 2025

Université de Lorraine, CHRU-Nancy, Department of Nuclear Medicine and Nancyclotep Imaging Platform, F-54000, Nancy, France; Université de Lorraine, INSERM U1254, IADI, F-54000 Nancy, France. Electronic address:

Background: This large-scale study analyzes factors affecting diagnostic accuracy of low-dose myocardial perfusion imaging and correlation with coronary angiography in a real-world practice.

Methods: We compared data extracted from routine reports of (i) low-dose [Tc]sestamibi stress-MPI performed with no attenuation correction and predominantly exercise stress testing and (ii) the corresponding coronary angiography.

Results: We considered 1070 pairs of coronary angiography/stress-MPI results reported by 11 physicians.

View Article and Find Full Text PDF
Article Synopsis
  • This study examined the effectiveness of bone SPECT imaging with and without SwiftScan technology in patients with prostate cancer bone metastases.
  • Involving 10 patients and 130 metastatic lesions, the research compared the quality of images produced at various acquisition times, showcasing improved quality scores and correlation coefficients with SwiftScan.
  • Findings indicated that SwiftScan allowed for a 50% reduction in acquisition time while still maintaining acceptable image quality and accurate quantitative measurements like SUV and contrast-to-noise ratios.
View Article and Find Full Text PDF

This study aimed to assess the diagnostic capability of quantitative parameters from whole-body bone single-photon emission computed tomography/computed tomography (SPECT/CT) in detecting bone metastases in prostate cancer patients; Methods: We retrospectively analyzed 82 prostate cancer patients who underwent staging bone scintigraphy with a full-ring 360° Cadmium-Zinc-Telluride (CZT) SPECT/CT system. From the SPECT/CT images, we measured the maximum (SUVmax) and mean (SUVmean) standardized uptake values at six normal bone sites (skull, humerus, thoracic spine, lumbar spine, iliac bone, and femur), and the SUVmax for both metastatic and benign bone lesions. Ratios of lesion SUVmax-to-maximum and mean uptake values at the skull, humerus, and femur were computed for each lesion; Results: SUVmax and SUVmean at the skull and femur exhibited significantly lower variance compared to those at the thoracic spine, lumbar spine, and iliac bone, and revealed no significant differences between patients with and without bone metastasis.

View Article and Find Full Text PDF

Recent Progress in Myocardial Perfusion Imaging Techniques.

Nucl Med Mol Imaging

December 2024

Department of Nuclear Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.

Article Synopsis
  • Nuclear cardiology has advanced rapidly, particularly in myocardial perfusion imaging (MPI) since the 1970s, which is vital for assessing and managing coronary artery disease.
  • Recent innovations include the development of gamma cameras with cadmium zinc telluride (CZT) systems and enhancements in software, as well as ongoing studies into new radiopharmaceuticals for positron emission tomography (PET)-MPI.
  • Despite these advancements, challenges like high equipment and examination costs, along with limited accessibility, pose significant barriers to the practical use of these technologies in nuclear medicine.
View Article and Find Full Text PDF

New Generation SPECT Cameras Based on Cadmium-Zinc Telluriide Technology.

Ann Nucl Cardiol

October 2024

Division of Nuclear Medicine, Department of Radiology, Saitama Medical University Hospital, Saitama, Japan.

Article Synopsis
  • Despite advances in positron emission tomography, cardiac SPECT imaging remains popular due to its accessibility and widespread clinical application.
  • CZT-based detectors offer improved imaging capabilities compared to traditional Anger type detectors, leading to successful clinical use since their introduction 15 years ago.
  • Ongoing research into new semiconductor materials and the development of advanced SPECT systems indicate potential future advancements in nuclear cardiology.
View Article and Find Full Text PDF