Objective: To evaluate low-tube-voltage 90-kVp CT pulmonary angiography (CTPA) with advanced modeled iterative reconstruction algorithm (Admire) compared to 120-kVp equivalent dual-energy (DE) acquisition with regards to radiation exposure, image quality and diagnostic accuracy for pulmonary embolism (PE) assessment.

Methods: CTPA studies of 40 patients with suspected PE (56.7 ± 16.3 years) performed on a third-generation 192-slice dual-source CT scanner were retrospectively included. 120-kVp equivalent linearly-blended (60% 90-kVp, 40% 150-kVp) and 90-kVp images were reconstructed. Attenuation and noise of the pulmonary trunk were measured to calculate contrast-to-noise ratios (CNR). Three radiologists assessed the presence of central and segmental PE and diagnostic confidence. Interobserver agreement was calculated using intraclass correlation coefficient (ICC). Radiation exposure was assessed as effective dose (ED).

Results: Pulmonary trunk CNR values were significantly increased in 90-kVp compared to linearly-blended series (15.4 ± 6.3 vs 11.3 ± 4.6, p < 0.001). Diagnostic accuracy for PE assessment was similar in both series with excellent interobserver agreement (p = 0.48; ICC, 0.83; p = 0.48). Overall confidence for PE assessment was rated excellent for both series with a significant advantage for linearly-blended series (p < 0.001; 4.1 vs 3.8). ED was reduced by 37.2% with 90-kVp compared to 120-kVp equivalent image series (1.1 ± 0.6 vs 1.7 ± 0.7 mSv, p < 0.001).

Conclusion: 90-kVp CTPA with Admire provided increased quantitative image quality with similar diagnostic accuracy and confidence for PE assessment compared to 120-kVp equivalent acquisition, while radiation dose was reduced by 37.2%. Advances in knowledge: 90-kVp CTPA with an advanced iterative reconstruction algorithm results in excellent image quality and reduction of radiation exposure without limiting diagnostic performance.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6209482PMC
http://dx.doi.org/10.1259/bjr.20180269DOI Listing

Publication Analysis

Top Keywords

image quality
16
diagnostic accuracy
16
120-kvp equivalent
16
iterative reconstruction
12
reconstruction algorithm
12
quality diagnostic
12
compared 120-kvp
12
radiation exposure
12
90-kvp
8
pulmonary angiography
8

Similar Publications

Background: Vestibular schwannoma (VS) is a common intracranial tumor that affects patients' quality of life. Reliable imaging techniques for tumor volume assessment are essential for guiding management decisions. The study aimed to compare the ABC/2 method to the gold standard planimetry method for volumetric assessment of VS.

View Article and Find Full Text PDF

Significance: Optimal meibography utilization and interpretation are hindered due to poor lid presentation, blurry images, or image artifacts and the challenges of applying clinical grading scales. These results, using the largest image dataset analyzed to date, demonstrate development of algorithms that provide standardized, real-time inference that addresses all of these limitations.

Purpose: This study aimed to develop and validate an algorithmic pipeline to automate and standardize meibomian gland absence assessment and interpretation.

View Article and Find Full Text PDF

Cardiac MR image reconstruction using cascaded hybrid dual domain deep learning framework.

PLoS One

January 2025

Medical Image Processing Research Group (MIPRG), Dept. of Elect. & Comp. Engineering, COMSATS University Islamabad, Islamabad, Pakistan.

Recovering diagnostic-quality cardiac MR images from highly under-sampled data is a current research focus, particularly in addressing cardiac and respiratory motion. Techniques such as Compressed Sensing (CS) and Parallel Imaging (pMRI) have been proposed to accelerate MRI data acquisition and improve image quality. However, these methods have limitations in high spatial-resolution applications, often resulting in blurring or residual artifacts.

View Article and Find Full Text PDF

Purpose: A relative afferent pupillary defect (RAPD) is a characteristic clinical sign of optic neuritis (ON). Here, we systematically evaluated ultrasound pupillometry (UP) for the detection of an RAPD in patients with ON, including a comparison with infrared video pupillometry (IVP), the gold standard for objective pupillometry.

Materials And Methods: We enrolled 40 patients with acute (n = 9) or past (n = 31) ON (ON+), 31 patients with multiple sclerosis (MS) without prior ON, and 50 healthy controls (HC) in a cross-sectional observational study.

View Article and Find Full Text PDF

Rationale: Solitary fibrous tumors (SFTs) are spindle cell tumors that typically occur in the pleura and peritoneum, but very rarely in the stomach. To our best knowledge, there are only 10 cases reported in English literature. We reported a case of primary stomach SFT and summarized the characteristics of all previous cases, suggesting that pathologists and surgeons should include this disease in the differential diagnosis list of primary mesenchymal tumor of the stomach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!