This paper describes a system for training healthcare practitioners in the identification of different arterial pulses. The driving system uses a linear solenoid in an open loop force control. Due to the large hysteresis it exhibited, a form of compensation was implemented, based on the classic Preisach model of hysteresis. Implementation of said model resulted in a significant reduction of force tracking error, demonstrating the feasibility of the chosen approach for the intended application.