Background: Conductive hearing loss due to ossicular abnormalities occurs from many causes, including trauma, infection, cholesteatoma, surgery and congenital anomalies. Surgical reconstruction of the ossicular chain is a well-established procedure for repair of ossicular defects, but is still plagued by high failure rates. Underlying disease and proper sizing of prostheses are two challenges that lead to component failure. Three-dimensional (3D) printing has been used successfully to solve a number of medical prosthesis problems. Custom 3D printing an individualized ossicular prosthesis would be a potential solution for the wide range of anatomic variation encountered in the pathological middle ear, and could decrease the rate of post-operative prosthesis displacement by increasing the likelihood of a proper fit, in addition to decreasing surgical time.In this study, the incus was removed from three formalin-fixed cadaveric human temporal bones with no macro- or microscopic evidence of pathology. Imaging of the cadaveric bone was obtained using a standard temporal bone CT protocol. A custom prosthesis for each cadaveric human temporal bone was designed using the Mimics Innovation Suite software (Materialise, Belgium) and fabricated on a Form2 3D printer (FormLabs, Somerville, Massachusetts). Four surgeons then performed insertion of each prosthesis into each middle ear, blinded to the bone from and for which each was designed. The surgeons were asked to match each prosthesis to its correct parent bone.

Results: Each prosthesis had unique measurements. Each of the four surgeons was able to correctly match the prosthesis model to its intended temporal bone. The chances of this occurring randomly are 1:1296.

Conclusions: A custom 3D printed ossicular prosthesis is a viable solution for conductive hearing loss due to ossicular chain defects. Commercially available CT scanners can detect significant anatomic differences in normal human middle ear ossicles. These differences can be accurately represented with current 3D printing technology and, more significantly, surgeons can detect these differences.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5954796PMC
http://dx.doi.org/10.1186/s41205-017-0015-2DOI Listing

Publication Analysis

Top Keywords

ossicular chain
12
ossicular prosthesis
12
middle ear
12
temporal bone
12
prosthesis
10
surgical reconstruction
8
ossicular
8
reconstruction ossicular
8
custom printed
8
printed ossicular
8

Similar Publications

This study aimed to investigate the efficacy of inverted door flap (IDF) combined with transcanal approach to the tympanic antrum (TCAA) technique under the endoscope for treatment of middle ear cholesteatoma. Outcomes of patients treated with combined techniques at the First People's Hospital of Foshan City between March 2021 and March 2023 were evaluated. A total of 31 patients (33 ears, 16 males and 15 females) aged (42.

View Article and Find Full Text PDF

Temporal bone CT is an essential technique for diagnosing ossicular chain trauma, and the location of standard observation planes (SOP) is the foundation of imaging diagnosis. The ossicular chain is small in volume, and there are about 11 standard observation planes for ossicular chain diagnosis, so it is a professional and time-consuming task to label SOPs accurately. An automatic annotation method of SOP is proposed.

View Article and Find Full Text PDF

BACKGROUND The Carhart effect consists of a reduction in bone conduction thresholds associated with conductive hearing loss. The aim of this study was to evaluate the role of the Carhart effect in predicting outcomes from surgery in 3 age groups. MATERIAL AND METHODS This study included 532 patients with conductive hearing loss due to otosclerosis, otitis media with effusion, and chronic otitis media who underwent surgery between 2010 and 2020.

View Article and Find Full Text PDF

Introduction: The tympanic cavity contains three tiny bones, the malleus, incus, and stapes, which have a fundamental role in the transmission of sound. Recent research emphasizes the use of CBCT for the anatomic study of the temporal bone. The information about middle ear anatomy on CBCT scans is meager; hence, this retrospective study was conducted to identify and determine the various morphometrical parameters of the malleus using CBCT which can be helpful during reconstructive procedures for the otologic surgeon.

View Article and Find Full Text PDF

Objective: The primary aim of this study was to investigate the accuracy of a semi-automatic algorithm in assessing the feasibility and complexity of endoscopic stapes surgery preoperatively.

Methods: A semi-automatic algorithm was developed to simulate endoscopic stapes surgery in 3D. To test the accuracy of the algorithm, five fresh-frozen cadaveric heads (ten ears) were used.

View Article and Find Full Text PDF