Purpose: We propose a novel compensator-based IMRT system designed to provide a simple, reliable, and cost-effective adjunct technology, with the goal of expanding global access to advanced radiotherapy techniques. The system would employ easily reusable tungsten bead compensators that operate independent of a gantry (e.g., mounted in a ring around the patient). Thereby the system can be retrofitted to existing linac and cobalt teletherapy units. This study explores the quality of treatment plans from the proposed system and the dependence on associated design parameters.

Methods: We considered Co-based plans as the most challenging scenario for dosimetry and benchmarked them against clinical MLC-based plans delivered on a linac. Treatment planning was performed in the Pinnacle treatment planning system with commissioning based on Monte Carlo simulations of compensated beams. Co-compensator IMRT plans were generated for five patients with head-and-neck cancer and five with gynecological cancer and compared to respective IMRT plans using a 6 MV linac beam with an MLC. The dependence of dosimetric endpoints on compensator resolution, thickness, position, and number of beams was assessed. Dosimetric accuracy was validated by Monte Carlo simulations of dose distribution in a water phantom from beams with the IMRT plan compensators.

Results: The Co-compensator plans had on average equivalent PTV coverage and somewhat inferior OAR sparing compared to the 6 MV-MLC plans, but the differences in dosimetric endpoints were clinically acceptable. Calculated treatment times for head-and-neck plans were 7.6 ± 2.0 min vs 3.9 ± 0.8 min (6 MV-MLC vs Co-compensator) and for gynecological plans were 8.7 ± 3.1 min vs 4.3 ± 0.4 min. Plan quality was insensitive to most design parameters over much of the ranges studied, with no degradation found when the compensator resolution was finer than 6 mm, maximum thickness at least 2 tenth-value-layers, and more than five beams were used. Source-to-compensator distances of 53 and 63 cm resulted in very similar plan quality. Monte Carlo simulations suggest no increase in surface dose for the geometries considered here. Simulated dosimetric validation tests had median gamma pass rates of 97.6% for criteria of 3% (global)/3 mm with a 10% threshold.

Conclusions: The novel ring-compensator IMRT system can produce plans of comparable quality to standard 6 MV-MLC systems. Even when Co beams are used the plan quality is acceptable and treatment times are substantially reduced. Co-compensator IMRT plans are adequately modeled in an existing commercial treatment planning system. These results motivate further development of this low-cost adaptable technology with translation through clinical trials and deployment to expand the reach of IMRT in low- and middle-income countries.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8522994PMC
http://dx.doi.org/10.1002/mp.12985DOI Listing

Publication Analysis

Top Keywords

treatment planning
16
imrt system
12
monte carlo
12
carlo simulations
12
imrt plans
12
plan quality
12
plans
11
imrt
8
system
8
low- middle-income
8

Similar Publications

Road traffic congestion on the cold chain logistics not only increase the cost and time, but also creates certain negative impact on the national carbon emissions. To fully utilize the traffic resources, this study has classified urban road traffic congestion and defined the various vehicle delivery speeds with dynamic congestion levels. Simultaneously, it has developed the cold chain products replenishment strategy by considering delivery route, multi-depot condition and even vehicle types, aiming to minimize the total cost and carbon emissions, and maximizing the cold chain products freshness.

View Article and Find Full Text PDF

Introduction: Death is universal, yet relatively little is known about how Canadians experience their death. Using novel decedent interview data from the Canadian Longitudinal Study on Aging we describe the prevalence and characteristics of peace with dying among older Canadians.

Methods: We conducted a secondary analysis of decedent interview data from the Canadian Longitudinal Study on Aging.

View Article and Find Full Text PDF

Background: Rifampin therapy is indicated for the treatment of staphylococcal periprosthetic joint infection (PJI) in patients who have undergone debridement, antibiotics, and implant retention (DAIR) or one-stage revision as per the Infectious Diseases Society of America (IDSA) guideline. Given the well-established effectiveness of rifampin as adjunctive therapy in staphylococcal PJI, it is crucial to evaluate its utilization in practice and identify factors that contribute to its underuse or incomplete administration, as these deviations may undermine treatment efficacy and patient outcomes.

Questions/purposes: Among patients who met clear indications for rifampin use having undergone DAIR or one-stage revision for staphylococcal PJI, (1) what proportion of patients did not receive it? (2) What proportion of patients started it but did not complete the planned course? (3) Where documented in the medical record, what were the common reasons for not using it or prematurely discontinuing it, and in what percentage of the patients' charts was no reason given? (4) What proportion of patients were taking a medication that put them at risk for a drug-drug interaction (DDI)?

Methods: Using an institutional database, patients who underwent DAIR or revision arthroplasty for PJI from January 2013 to April 2023 were identified (n = 935).

View Article and Find Full Text PDF

Optimal management of adult congenital heart disease (ACHD) requires a multidisciplinary team (MDT) approach, fostering a collaborative culture over an individualistic approach. Within this framework, subspecialty-trained radiologists provide crucial imaging expertise, supporting cardiologists, surgeons, and interventional cardiologists in diagnoses, treatment planning, and follow-up evaluations. Advanced imaging tools and a nuanced understanding of surgical and interventional procedures enable radiologists to provide valuable insights to clinicians.

View Article and Find Full Text PDF

With the increasing prevalence of knee osteoarthritis (KOA), the limitations of traditional treatments, such as their limited efficacy in halting disease progression and their potential side effects, are becoming more evident. This situation has prompted scientists to seek more effective strategies. In recent years, exercise therapy has gained prominence in KOA treatment due to its safety, efficacy, and cost-effectiveness, which are underpinned by the molecular actions of exerkines.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!