The effect of light intensity on image quality in endoscopic ear surgery.

Clin Otolaryngol

Department of Otolaryngology, Monklands Hospital, Airdrie, Scotland.

Published: October 2018

Objectives: Endoscopic ear surgery is a rapidly developing field with many advantages. But endoscopes can reach temperatures of over 110°C at the tip, raising safety concerns. Reducing the intensity of the light source reduces temperatures produced. However, quality of images at lower light intensities has not yet been studied. We set out to study the effect of light intensity on image quality in EES.

Design: Prospective study of patients undergoing EES from April to October 2016. Consecutive images of the same operative field at 10%, 30%, 50% and 100% light intensities were taken. Eight international experts were asked to each evaluate 100 anonymised, randomised images.

Setting: District General Hospital.

Participants: Twenty patients.

Main Outcome Measures: Images were evaluated on a 5-point Likert scale (1 = significantly worse than average; 5 = significantly better than average) for detail of anatomy; colour contrast; overall quality; and suitability for operating.

Results: Mean scores for photographs at 10%, 30%, 50% and 100% light intensity were 3.22 (SD 0.93), 3.15 (SD 0.84), 3.08 (SD 0.88) and 3.10 (SD 0.86), respectively. In ANOVA models for the scores on each of the scales (anatomy, colour contrast, overall quality and suitability for operating), the effects of rater and patient were highly significant (P < .0005) but light intensity was non-significant (P = .34, .32, .21, .15, respectively).

Conclusion: Images taken during surgery by our endoscope and operative camera have no loss of quality when taken at lower light intensities. We recommend the surgeon considers use of lower light intensities in endoscopic ear surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1111/coa.13139DOI Listing

Publication Analysis

Top Keywords

light intensity
12
intensity image
8
image quality
8
endoscopic ear
8
ear surgery
8
light intensities
8
10% 30%
8
30% 50%
8
50% 100%
8
100% light
8

Similar Publications

The endo-lysosomal system plays a crucial role in maintaining cellular homeostasis and promoting organism fitness. The pH of its acidic compartments is a crucial parameter for proper function, and it is dynamically influenced by both intracellular and environmental factors. Here, we present a method based on fluorescence lifetime imaging microscopy (FLIM) for quantitatively analyzing the pH profiles of acidic endolysosomal compartments in diverse types of primary mammalian cells and in live organism .

View Article and Find Full Text PDF

Chlorin e6: a promising photosensitizer of anti-tumor and anti-inflammatory effects in PDT.

Nanomedicine (Lond)

January 2025

Jiangsu Key Laboratory of New Drug Research and Clinical Pharmacy, Xuzhou Medical University, Xuzhou, China.

Photodynamic therapy (PDT) involves the activation of photosensitizers (PSs) by visible laser light at the target site to catalyze the production of reactive oxygen species, resulting in tumor cell death and blood vessel closure. The efficacy of PDT depends on the PSs, the amount of oxygen, and the intensity of the excitation laser. PSs have been extensively researched, and great efforts have been made to develop an ideal photosensitizer.

View Article and Find Full Text PDF

Characterization and application of fluorescent hydrogel films with superior mechanical properties in detecting iron(Ⅲ) ions and ferroptosis in oral cancer.

Front Bioeng Biotechnol

January 2025

Department of Orthodontics, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, College of Stomatology, Shanghai Jiao Tong University, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Shanghai Key Laboratory of Stomatology and Shanghai Research Institute of Stomatology, Research Unit of Oral and Maxillofacial Regenerative Medicine, Chinese Academy of Medical Sciences, Shanghai, China.

A one-step hydrothermal method was applied to prepare carbon dots (CDs) with superior fluorescence properties using chitosan as a carbon source. The as-prepared carbon dots were then grafted onto a sodium alginate-gelatin hydrogel film to form a fluorescent hydrogel film (FHGF), emitting at 450 nm under excitation of 350 nm light. In comparison to the CDs, the fluorescence intensity of this film was maintained over 90.

View Article and Find Full Text PDF

Metasurfaces consisting of subwavelength structures have shown unparalleled capability in light field manipulation. However, their functionalities are typically static after fabrication, limiting their practical applications. Though persistent efforts have led to dynamic wavefront control with various materials and mechanisms, most of them work in free space and require specialized materials or bulky configurations for external control.

View Article and Find Full Text PDF

Optoelectronic tweezers (OET) offer a versatile, programmable, and contactless method for manipulating microscale objects. While factors like AC voltage and light intensity have been extensively studied, the role of light pattern curvature in the performance of OET manipulation remains underexplored. This study investigates how the curvature of light patterns affects the movement of polystyrene microparticles under negative dielectrophoretic (DEP) forces in an OET system.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!