Objectives: This study sought to evaluate basket catheter deployment, catheter-tissue contact, and time-space stability of unipolar atrial electrograms (aEGMs) recorded in persistent atrial fibrillation (AF) patients.

Background: Panoramic mapping of human AF using multiple-electrode basket catheters may identify AF sources. Although clinical results using this technique are provocative, questions remain about its effectiveness.

Methods: Data were collected from patients (N = 25) undergoing catheter ablation for AF during the multicenter STARLIGHT (Signal Transfer of Atrial Fibrillation Data to Guide Human Treatment) trial (NCT01765075). Left and right aEGM signals were recorded using basket catheters during baseline AF, following ablation and during sinus rhythm. Data were analyzed for basket deployment, peak-to-peak voltage, and electrogram stability and organization. Electrogram stability and organization were evaluated via time-frequency analysis (TFA).

Results: Basket catheters displayed equatorial bunching when deployed in atria. Interspline spacing ranged from 1.7 to 64.0 mm in the right atrial and from 1.5 to 85.08 mm in the left atrial basket. Approximately one-third of mapping electrodes failed to demonstrate a median peak-to-peak voltage >2× the low-voltage threshold. Time-space stability and organization was observed in 13 of 22 (59.09%) right atrial and 10 of 22 (45.45%) left atrial baskets.

Conclusions: Despite poor deployment and a large number of low-voltage electrodes, stability and organization was observed in about one-half of the mapped patients. Although this study suggests that basket catheters have limitations for patient-specific AF mapping, concordant activation occurs in some persistent AF patients, which may be amenable to high-density mapping techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jacep.2015.09.017DOI Listing

Publication Analysis

Top Keywords

basket catheters
20
stability organization
16
atrial fibrillation
12
atrial
8
persistent atrial
8
basket
8
time-space stability
8
peak-to-peak voltage
8
electrogram stability
8
left atrial
8

Similar Publications

Background: Electrographic flow (EGF) mapping allows for the visualization of global atrial wavefront propagations. One mechanism of initiation and maintenance of atrial fibrillation (AF) is stimulation from EGF-identified focal sources that serve as driver sites of fibrillatory conduction. Electrographic flow consistency (EGFC) further quantifies the concordance of observed wavefront patterns, indicating that a healthier substrate shows more organized wavefront propagation and higher EGFC.

View Article and Find Full Text PDF

Preliminary validation for an online configuration determination method of a thin film buckling under point contact force.

Sci Rep

December 2024

Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea.

The authors previously developed an online thin film buckling configuration determination method for a mini basket type mapping catheter prototype, which incorporates eight thin film sensor strips. In the prior study, no external force was applied to the thin film, and only axial displacement was adjusted to induce buckling in the thin film. Extending this prior work, a preliminary methodological validation is conducted for an online configuration determination method of thin film buckling under a point contact force.

View Article and Find Full Text PDF
Article Synopsis
  • Stent dislodgement is a rare but difficult complication during heart procedures, and existing treatment methods are not very effective.
  • This study tested a new self-expanding basket (SEB) catheter designed to retrieve dislodged stents without causing additional injury or needing surgery.
  • In a small trial with 6 patients, the SEB catheter successfully retrieved all dislodged stents, with no safety issues reported, indicating it may be a promising solution, but more extensive research is needed.
View Article and Find Full Text PDF

Lesion Morphometry of the Pentaspline Pulsed Field Ablation Catheter: Understanding Catheter Pose, Rotation, and Dosing.

Circ Arrhythm Electrophysiol

December 2024

Helmsley Electrophysiology Center, Department of Cardiology, Icahn School of Medicine at Mount Sinai, New York, NY (K.W., M.N., V.Y.R., J.S.K.).

Background: The pentaspline pulsed field ablation catheter achieves pulmonary vein isolation using 8 stacked, pose-specific applications with rotation. The morphology of pose-specific, single or double applications has not been described.

Methods: One or 2 applications were delivered to select veins and discrete atrial and ventricular sites in 9 swine.

View Article and Find Full Text PDF
Article Synopsis
  • Electrographic flow (EGF) mapping is a technique used to understand how electrical signals move in the heart's atria, especially in relation to atrial fibrillation (AF) compared to normal sinus rhythm (SR).
  • The study involved mapping electrical activity in the atria of ten patients, showing that EGF patterns differ significantly in AF, with lower electrographic flow consistency (EGFC) during AF compared to SR, indicating more chaotic activity.
  • The findings suggest EGFC could serve as a valuable marker for assessing atrial health and myopathy, correlating with bipolar voltage measurements, which are commonly used to evaluate heart conditions.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!