Introduction: Cholangiopancreatoscopy with SpyGlass™ Direct Visualization System (SGDVS) is being used in specialized centers for improving the sensitivity of endoscopic retrograde cholangiopancreatography (ERCP) in patients with indeterminate pancreatobiliary strictures (PBS). The aims of this study were to report our initial experience with SGDVS in the evaluation of indeterminate PBS, and discuss the improvements of ERCP brought by this technique in our center.
Methods: The usefulness of SGDVS in patients with indeterminate PBS (defined after nondiagnostic previous ERCP with brush cytology) was evaluated in a prospective observational cohort study conducted at a single tertiary biliopancreatic unit. The accuracy of diagnosis by the SGDVS visual findings, SGDVS-guided biopsy, technical success, image quality, change in patient management after the procedure, and complication rate were assessed.
Results: In our single-center cohort, there were 13 SGDVS procedures for evaluating indeterminate PBS. Technical success, defined by the ability to progress with the SpyScope to the target lesion, was achieved in all the cases. The diagnostic accuracy of visual findings (87.5%) was superior to SGDVS-guided biopsy (55%). In 11 (85%) procedures, the image quality was considered good. The procedure permitted exclusion of malignancy and avoiding surgery in 9 patients (69%). There were no complications during the procedures. However, in the post-procedure monitoring, 3 patients developed acute pancreatitis (19%) and 2 patients developed acute cholangitis (13%).
Conclusion: The SGDVS can be considered useful in the context of indeterminate PBS. The intervention is associated with high procedural success and alters clinical outcome compared to conventional approaches.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5939843 | PMC |
http://dx.doi.org/10.1159/000481859 | DOI Listing |
PLoS One
April 2023
Infectious Diseases Research Collaboration, Kampala, Uganda.
Background: Stool is a potential sample for diagnosing Mycobacterium tuberculosis (Mtb) in patients with difficulty in expectorating. However, high mycobacterial culture contamination rates and Xpert MTB/RIF Ultra test error rates on stool samples have limited its use. OMNIgene SPUTUM (OM-S) is a sample transport reagent with characteristics of sputum decontamination while maintaining viable Mtb.
View Article and Find Full Text PDFPract Lab Med
November 2019
Department of Pathology, University of Utah, 500 Chipeta Way, Salt Lake City, UT, 84108, USA.
Objectives: Competitive immunoenyzmatic assays for estradiol (E2) and unconjugated estriol (uE3) on UniCel DxI 800 Access immunoassay systems (Beckman Coulter) utilize bovine alkaline phosphatase (ALP) for amplification. In these assays, rare 'IND' error flags indicate that a relative light unit (RLU) raw result is past the high or low end of the calibration curve but cannot be differentiated from an instrument error or analytical interference. The present studies were conducted to establish a protocol to identify analytical interference and to characterize its mechanism when present.
View Article and Find Full Text PDFStroke
January 2019
From the Institute of Clinical Radiology (M.B., T.N., W.H., M.W., A.K., P.B.S.), Westfaelische Wilhelms-University of Muenster and Albert-Schweitzer-Campus 1, Germany.
Background and Purpose- Arterial ischemic stroke (AIS) in childhood is a severe disease with potentially lifelong restrictions. Apart from cardiac or prothrombotic embolism, arteriopathy has been identified as a major cause and significant target of secondary stroke prevention. The Childhood Arterial Ischemic Stroke Standardized Classification and Diagnostic Evaluation (CASCADE) criteria facilitate a structured tool for categorizing and reporting childhood strokes according to the underlying cause.
View Article and Find Full Text PDFGE Port J Gastroenterol
April 2018
Gastroenterology Department - Centro Hospitalar de São João, WGO Oporto Training Center, Porto Medical School - University of Porto, Porto, Portugal.
Introduction: Cholangiopancreatoscopy with SpyGlass™ Direct Visualization System (SGDVS) is being used in specialized centers for improving the sensitivity of endoscopic retrograde cholangiopancreatography (ERCP) in patients with indeterminate pancreatobiliary strictures (PBS). The aims of this study were to report our initial experience with SGDVS in the evaluation of indeterminate PBS, and discuss the improvements of ERCP brought by this technique in our center.
Methods: The usefulness of SGDVS in patients with indeterminate PBS (defined after nondiagnostic previous ERCP with brush cytology) was evaluated in a prospective observational cohort study conducted at a single tertiary biliopancreatic unit.
Aim: to assess the inter observer reproducibility of Single Photon Computed Tomography-Computerised Tomography (SPECT-CT) in classifying indeterminate foci of abnormal metabolism on planar bone scan (PBS), in oncologic diseases, in the limit of performance of a low dose CT.
Methods: 49 patients (27 men, 22 women) whose PBS showed indeterminate lesions underwent SPECT/CT. The mean age was 57 ± 13 years (range 22-80 years).