Objectives: The primary objective was to assess the acute and 3-month performance of the modular antitachycardia pacing (ATP)-enabled leadless pacemaker (LP) and subcutaneous implantable cardioverter-defibrillator (S-ICD) system, particularly device-device communication and ATP delivery.

Background: Transvenous pacemakers and implantable cardioverter-defibrillators (ICDs) have considerable rates of lead complications. We examined the next step in multicomponent leadless cardiac rhythm management: feasibility of pacing (including ATP) by a LP, commanded by an implanted S-ICD through wireless, intrabody, device-device communication.

Methods: The combined modular cardiac rhythm management therapy system of the LP and S-ICD prototypes was evaluated in 3 animal models (ovine, porcine, and canine) both in acute and chronic (90 days) experiments. LP performance, S-ICD to LP communication, S-ICD and LP rhythm discrimination, and ATP delivery triggered by the S-ICD were tested.

Results: The LP and S-ICD were successfully implanted in 98% of the animals (39 of 40). Of the 39 animals, 23 were followed up for 90 days post-implant. LP performance was adequate and exhibited appropriate VVI behavior during the 90 days of follow-up in all tested animals. Unidirectional communication between the S-ICD and LP was successful in 99% (398 of 401) of attempts, resulting in 100% ATP delivery by the LP (10 beats at 81% of the coupling interval). Adequate S-ICD sensing was observed during normal sinus rhythm, LP pacing, and ventricular tachycardia/ventricular fibrillation.

Conclusions: This study presents the preclinical acute and chronic performance of the combined function of an ATP-enabled LP and S-ICD. Appropriate VVI functionality, successful wireless device-device communication, and ATP delivery were demonstrated by the LP. Clinical studies on safety and performance are needed.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jacep.2017.04.002DOI Listing

Publication Analysis

Top Keywords

atp delivery
12
s-icd
10
acute 3-month
8
3-month performance
8
pacemaker subcutaneous
8
subcutaneous implantable
8
device-device communication
8
communication atp
8
cardiac rhythm
8
rhythm management
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!