AI Article Synopsis

  • The study assessed the accuracy of measuring implant lengths using two different image-viewers on CBCT images and looked at how artifacts from implants might affect these measurements.
  • A resin model of an edentulous mandible with several implants was used, allowing two observers to independently measure implant lengths across different viewing software.
  • Results showed no significant differences between the actual and measured lengths, indicating that both image-viewers provided comparable accuracy and that surrounding artifacts from the implants did not skew the measurements.

Article Abstract

This preclinical study compared the accuracy of implant lengths measured in two different image-viewers, and examined whether implant-induced artifacts affected the implant length measurements on CBCT images. A resin edentulous mandibular model, with multiple adjacent implants in the posterior segments, was acquired with a CBCT machine. In two different image-viewers, two observers independently measured the implant length. Vertical measurements on CBCT images were carried out twice at each session, and repeated one week later. The results demonstrated no significant differences between actual and measured implant lengths. The differences in the mean error for vertical measurements using the two different image-viewers (cross-sectional images: OsiriX viewer = -0.01 ± 0.03 mm, NewTom viewer = -0.05 ± 0.09 mm, -value = 0.056; sagittal images: OsiriX viewer = -0.03 ± 0.04 mm; NewTom viewer = -0.04 ± 0.10 mm, -value = 0.24) were not statistically significant. This investigation suggests that the accuracy of implant length measurements on CBCT images was not influenced by image-viewers or by the presence of implant-induced artifacts. The presence of multiple adjacent implants in the posterior segments of the mandible is not likely to impact the measurements made between the implant apex and vital structures on CBCT images.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5839189PMC
http://dx.doi.org/10.1002/cre2.18DOI Listing

Publication Analysis

Top Keywords

implant length
16
cbct images
16
length measurements
12
measurements cbct
12
artifacts implant
8
accuracy implant
8
implant lengths
8
implant-induced artifacts
8
multiple adjacent
8
adjacent implants
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!