Objectives: In an attempt to improve spectral resolution and speech intelligibility, several current focusing methods have been proposed to increase spatial selectivity by decreasing intracochlear current spread. For example, tripolar stimulation administers current to a central electrode and uses the two flanking electrodes as the return pathway, creating a narrower intracochlear electrical field and hence increases spectral resolution when compared with monopolar (MP) stimulation. However, more current is required, and in some patients, specifically the ones with high electrode impedances, full loudness growth cannot be supported because of compliance limits. The present study describes and analyses a new loudness encoding approach that uses tripolar stimulation near threshold and gradually broadens the excitation (by decreasing compensation coefficient σ) to increase loudness without the need to increase overall current. It is hypothesized that this dynamic current focusing (DCF) strategy increases spatial selectivity, especially at lower loudness levels, while maintaining maximum selectivity at higher loudness levels, without reaching compliance limits.

Design: Eleven adult cochlear implant recipients with postlingual hearing loss, with at least 9 months of experience with their HiRes90K implant, were selected to participate in this study. Baseline performance regarding speech intelligibility in noise (Dutch matrix sentence test), spectral ripple discrimination at 45 and 65 dB, and temporal modulation detection thresholds were assessed using their own clinical program, fitted on a Harmony processor. Subsequently, the DCF strategy was fitted on a research Harmony processor. Threshold levels were determined with σ = 0.8, which means 80% of current is returned to the flanking electrodes and the remaining 20% to the extracochlear ground electrode. Instead of increasing overall pulse magnitude, σ was decreased to determine most comfortable loudness. After 2 to 3 hr of adaptation to the research strategy, the same psychophysical measures were taken.

Results: At 45 dB, average spectral ripple scores improved significantly from 2.4 ripples per octave with their clinical program to 3.74 ripples per octave with the DCF strategy (p = 0.016). Eight out of 11 participants had an improved spectral resolution at 65 dB. Nevertheless, no significant difference between DCF and MP was observed at higher presentation levels. Both speech-in-noise and temporal modulation detection thresholds were equal for MP and DCF strategies. Subjectively, 2 participants preferred the DCF strategy over their own clinical program, 2 preferred their own strategy, while the majority of the participants had no preference. Battery life was decreased and ranged from 1.5 to 4 hr.

Conclusions: The DCF strategy gives better spectral resolution, at lower loudness levels, but equal performance on speech tests. These outcomes warrant for a longer adaptation period to study long-term outcomes and evaluate if the outcomes in the ripple tests transfer to the speech scores. Further research, for example, with respect to fitting rules and reduction of power consumption, is necessary to make the DCF strategy suitable for routine clinical application.

Download full-text PDF

Source
http://dx.doi.org/10.1097/AUD.0000000000000593DOI Listing

Publication Analysis

Top Keywords

dcf strategy
24
spectral resolution
16
current focusing
12
loudness levels
12
clinical program
12
dynamic current
8
loudness
8
speech intelligibility
8
spatial selectivity
8
tripolar stimulation
8

Similar Publications

Low friction hydrogel with diclofenac eluting ability for dry eye therapeutic contact lenses.

Methods

November 2024

Centro de Química Estrutural (CQE), Institute of Molecular Sciences, Departamento de Engenharia Química, Instituto Superior Técnico, Universidade de Lisboa, Av. Rovisco Pais, 1049-001 Lisbon, Portugal; Egas Moniz Center for Interdisciplinary Research (CiiEM), Egas Moniz School of Health & Science, 2829-511 Caparica, Almada, Portugal. Electronic address:

When placed in the eye, contact lenses (CLs) disturb the tear fluid and affect the natural tribological behaviour of the eye. The disruption in the contact mechanics between the ocular tissues can increase frictional shear stress and ocular dryness, causing discomfort. Ultimately, continuous CLs wear can trigger inflammation which is particularly critical for people suffering from dry eye.

View Article and Find Full Text PDF

Exposure to 2,4-Dichlorophenoxyacetic acid stimulates the calcium/ROS/CK1α pathway to trigger Hemolysis and Eryptosis in red blood cells.

Toxicol Res (Camb)

December 2024

Department of Clinical Laboratory Sciences, College of Applied Medical Sciences, King Saud University, King Khalid Road, Riyadh 12372, Saudi Arabia.

Objective: The agricultural herbicide 2,4-dichlorophenoxyacetic (2,4-D) is cytotoxic to human red blood cells (RBCs) by virtue of oxidative hemolysis. Nevertheless, there remains a great paucity of literature detailing the mechanisms by which 2,4-D triggers hemolysis. Also, the eryptotic effects of 2,4-D has thus far been largely overlooked.

View Article and Find Full Text PDF

Introduction: Despite the progress in gene editing platforms like CRISPR/Cas9 with the potential to transform the standard of care for haemophilia, the language used to explain and discuss gene editing is not aligned across the haemophilia community. Here, we present the objective and rationale for developing a clear, consistent, and globally aligned gene editing lexicon to address these communication gaps.

Methods: Effectively communicating complex gene editing concepts requires a clear and consistent vocabulary.

View Article and Find Full Text PDF
Article Synopsis
  • * Porous silicon nanoparticles can effectively encapsulate APIs in an amorphous state, enhancing their dissolution rates compared to crystalline forms, and a new synthesis method using centrifugal Chemical Vapor Deposition (cCVD) has been developed to facilitate this process.
  • * Research showed that cCVD Si particles significantly increased dissolution rates for various model drugs, with additional improvements when drug–cyclodextrin complexes were loaded, suggesting effective strategies for enhancing drug bioavailability.
View Article and Find Full Text PDF

Seeking highly efficient adsorbents for pharmaceuticals and personal care products (PPCPs) removal has been a worldwide continuing endeavor. In this study, a new 3D composite material was synthesized by covalently anchoring Poly(m-Phenylenediamine) onto 3D polyvinyl alcohol modified foam framework (PmPD-MF-PVA). PmPD-MF-PVA was characterized and evaluated for its efficacy in removing diclofenac (DCF), a commonly detected PPCPs in both wastewater and surface water.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!