Fracture loads and failure modes of customized and non-customized zirconia abutments.

Dent Mater

Department of Dental Materials and Prosthodontics, Dental School of Ribeirão Preto, University of São Paulo, Avenida do Café, s/n, Monte Alegre, Ribeirão Preto, SP CEP 14040-904, Brazil. Electronic address:

Published: August 2018

Objective: This study aimed to evaluate the fracture load and pattern of customized and non-customized zirconia abutments with Morse-taper connection.

Methods: 18 implants were divided into 3 groups according to the abutments used: Zr - with non-customized zirconia abutments; Zrc - with customized zirconia abutments; and Ti - with titanium abutments. To test their load capacity, a universal test machine with a 500-kgf load cell and a 0.5-mm/min speed were used. After, one implant-abutment assembly from each group was analyzed by Scanning Electron Microscopy (SEM). For fractographic analysis, the specimens were transversely sectioned above the threads of the abutment screw in order to examine their fracture surfaces using SEM.

Results: A significant difference was noted between the groups (Zr=573.7±11.66N, Zrc=768.0±8.72N and Ti=659.1±7.70N). Also, the zirconia abutments fractured while the titanium abutments deformed plastically. Zrc presented fracture loads significantly higher than Zr (p=0.009). All the zirconia abutments fractured below the implant platform, starting from the area of contact between the abutment and implant and propagating to the internal surface of the abutment. All the zirconia abutments presented complete cleavage in the mechanical test. Fractography detected differences in the position and pattern of fracture between the two groups with zirconia abutments, probably because of the different diameters in the transmucosal region.

Significance: Customization of zirconia abutments did not affect their fracture loads, which were comparable to that of titanium and much higher than the maximum physiological limit for the anterior region of the maxilla.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.dental.2018.04.005DOI Listing

Publication Analysis

Top Keywords

zirconia abutments
36
fracture loads
12
non-customized zirconia
12
abutments
12
zirconia
9
customized non-customized
8
titanium abutments
8
abutments fractured
8
fracture
6
loads failure
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!