Orbital Reconstruction by Patient-Specific Implant Printed in Porous Titanium: A Retrospective Case Series of 12 Patients.

J Oral Maxillofac Surg

Associate Professor, Aix Marseille Univeristy, APHM, IFSTTAR, LBA, CHU Nord et Conception, Service de chirurgie maxillo-faciale, Marseille, France.

Published: October 2018

Purpose: The purpose of this study was to evaluate the orbital patient-specific implant (PSI) directly printed in porous titanium for the reconstruction of complex orbital bone defects in a series of 12 patients.

Patients And Methods: The authors designed and implemented a case series. The sample consisted of patients with unilateral complex orbital bone loss. All patients received a porous titanium PSI designed from the healthy contralateral side (mirroring). The criteria analyzed were the functional results: correction of enophthalmos, correction of ocular motility, operative time, complications, and operative revisions. The study was performed from 2015 through 2017.

Results: The sample was composed of 12 patients (mean age, 47 yr; age range, 13 to 70 yr). Patients were followed for a mean of 36 weeks postoperatively (range, 4 to 100 weeks). Twelve of the 12 patients presented preoperative enophthalmia, and 8 of the 12 patients presented preoperative diplopia. The mean operating time was 71 minutes (range, 60 to 200 minutes). For 8 patients, the follow-up was simple. In contrast, 2 patients required surgical revision with repositioning of the implant because of intraoperative implant malpositioning with esthetic or functional disturbance and malpositioning was confirmed on the postoperative computed tomogram, 1 patient required explantation of his implant 7 months after the surgery because of spheno-orbital meningioma recurrence (the implant was well positioned), and 1 patient operated on by a subciliary approach presented a postoperative ectropion. In this series of porous titanium orbital PSIs without positioning guides, 17% had malpositioning (2 patients who required a new intervention for repositioning).

Conclusion: The results of this study suggest that porous titanium PSI could be a surgical option for patients with complex orbital bone defects. In this series 17% of the sample needed a second operation. There are several ways to improve these results, such as intraoperative navigation or integrated positioning guides.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.joms.2018.04.006DOI Listing

Publication Analysis

Top Keywords

porous titanium
20
complex orbital
12
orbital bone
12
patients
11
patient-specific implant
8
printed porous
8
case series
8
bone defects
8
defects series
8
titanium psi
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!