The aim of the study was to investigate patient-related factors associated with either reliable or poorly reliable measurement results of ultrasound-based shear wave elastography (SWE) of the liver. A total of 188 patients were analyzed prospectively with binary logistic regression using the interquartile range/median as cutoff to define two groups based on reliable and poorly reliable SWE results. SWE results correlated significantly with liver biopsy. Factors associated with reliable SWE results (i.e., no negative impact on measurements) were age, sex, cirrhosis, antiviral and/or cardiovascular medication, smoking habits and body mass index. Factors associated with poorly reliable SWE results were increased skin-to-liver capsule distance (odds ratio = 3.08, 95% confidence interval: 1.70-5.60) and steatosis (odds ratio = 2.89, 95% confidence interval: 1.33-6.28). These findings indicate that the interquartile range/median as a quality parameter is useful in avoiding poorly reliable SWE results. How best to examine patients with increased skin-to-liver capsule distance is a matter of some controversy, as the incidences of obesity, diabetes and metabolic syndrome are increasing worldwide; however, our results indicate that reliable SWE results can be obtained in this group of patients by using ultrasound-based SWE.