The objective of this study was to develop instant thin-layer chromatography (ITLC) conditions for the determination of radiochemical purity of Ga-DOTATATE in a shorter time period than those stated in the NETSPOT (Advanced Accelerator Applications, Saint-Genis-Pouilly, France; AAA) kit package insert (PI). A faster ITLC system is needed to reduce the 48- to 50-min development time so that more radioactivity is available for single patient use and wait times are shorter in the event of kit failure. Variations of the PI mobile system were evaluated with microfiber chromatography paper impregnated with silica gel (ITLC-SG). After a more suitable mobile system was identified, evaluation began by attempting to shorten the 10-cm development distance to 7, 8, and 9 cm. Experiments using variations of PI mobile phase showed that increasing the proportion of methanol in the mobile phase decreased development time. Additionally, if the ratio of 1 M ammonium acetate was reduced to 10% or less, retention factor values fall outside specification. Reducing the development distance shortened development time as expected; however, it also affected the resolution aspect of the radiochromatogram. The fastest developing ITLC system, which maintained resolution and peak shape, was methanol:1 M ammonium acetate (80:20 V/V) with ITLC-SG using a development distance of 8 cm.