Background: Improving the signal-to-noise ratio (SNR) of chemical-shift-encoded MRI acquisition with complex reconstruction (MRI-C) may improve the accuracy and precision of noninvasive proton density fat fraction (PDFF) quantification in patients with hepatic steatosis.
Purpose: To assess the accuracy of high SNR (Hi-SNR) MRI-C versus standard MRI-C acquisition to estimate hepatic PDFF in adult and pediatric nonalcoholic fatty liver disease (NAFLD) using an MR spectroscopy (MRS) sequence as the reference standard.
Study Type: Prospective.
Population/subjects: In all, 231 adult and pediatric patients with known or suspected NAFLD.
Field Strength/sequence: PDFF estimated at 3T by three MR techniques: standard MRI-C; a Hi-SNR MRI-C variant with increased slice thickness, decreased matrix size, and no parallel imaging; and MRS (reference standard).
Assessment: MRI-PDFF was measured by image analysts using a region of interest coregistered with the MRS-PDFF voxel.
Statistical Tests: Linear regression analyses were used to assess accuracy and precision of MRI-estimated PDFF for MRS-PDFF as a function of MRI-PDFF using the standard and Hi-SNR MRI-C for all patients and for patients with MRS-PDFF <10%.
Results: In all, 271 exams from 231 patients were included (mean MRS-PDFF: 12.6% [SD: 10.4]; range: 0.9-41.9). High agreement between MRI-PDFF and MRS-PDFF was demonstrated across the overall range of PDFF, with a regression slope of 1.035 for the standard MRI-C and 1.008 for Hi-SNR MRI-C. Hi-SNR MRI-C, compared to standard MRI-C, provided small but statistically significant improvements in the slope (respectively, 1.008 vs. 1.035, P = 0.004) and mean bias (0.412 vs. 0.673, P < 0.0001) overall. In the low-fat patients only, Hi-SNR MRI-C provided improvements in the slope (1.058 vs. 1.190, P = 0.002), mean bias (0.168 vs. 0.368, P = 0.007), intercept (-0.153 vs. -0.796, P < 0.0001), and borderline improvement in the R (0.888 vs. 0.813, P = 0.01).
Data Conclusion: Compared to standard MRI-C, Hi-SNR MRI-C provides slightly higher MRI-PDFF estimation accuracy across the overall range of PDFF and improves both accuracy and precision in the low PDFF range.
Level Of Evidence: 1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2019;49:229-238.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6205914 | PMC |
http://dx.doi.org/10.1002/jmri.26168 | DOI Listing |
Abdom Radiol (NY)
October 2019
Liver Imaging Group, Department of Radiology, University of California San Diego, San Diego, CA, USA.
Purpose: This study compares splenic proton density fat fraction (PDFF) measured using confounder-corrected chemical shift-encoded (CSE)-MRI to magnetic resonance spectroscopy (MRS) in human patients at 3T.
Methods: This was a prospectively designed ancillary study to various previously described single-center studies performed in adults and children with known or suspected nonalcoholic fatty liver disease. Patients underwent magnitude-based MRI (MRI-M), complex-based MRI (MRI-C), high signal-to-noise variants (Hi-SNR MRI-M and Hi-SNR MRI-C), and MRS at 3T for spleen PDFF estimation.
J Magn Reson Imaging
January 2019
Liver Imaging Group, Department of Radiology, University of California at San Diego, La Jolla, California, USA.
Background: Improving the signal-to-noise ratio (SNR) of chemical-shift-encoded MRI acquisition with complex reconstruction (MRI-C) may improve the accuracy and precision of noninvasive proton density fat fraction (PDFF) quantification in patients with hepatic steatosis.
Purpose: To assess the accuracy of high SNR (Hi-SNR) MRI-C versus standard MRI-C acquisition to estimate hepatic PDFF in adult and pediatric nonalcoholic fatty liver disease (NAFLD) using an MR spectroscopy (MRS) sequence as the reference standard.
Study Type: Prospective.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!