Purpose: The task-based assessment of image quality using model observers is increasingly used for the assessment of different imaging modalities. However, the performance computation of model observers needs standardization as well as a well-established trust in its implementation methodology and uncertainty estimation. The purpose of this work was to determine the degree of equivalence of the channelized Hotelling observer performance and uncertainty estimation using an intercomparison exercise.

Materials And Methods: Image samples to estimate model observer performance for detection tasks were generated from two-dimensional CT image slices of a uniform water phantom. A common set of images was sent to participating laboratories to perform and document the following tasks: (a) estimate the detectability index of a well-defined CHO and its uncertainty in three conditions involving different sized targets all at the same dose, and (b) apply this CHO to an image set where ground truth was unknown to participants (lower image dose). In addition, and on an optional basis, we asked the participating laboratories to (c) estimate the performance of real human observers from a psychophysical experiment of their choice. Each of the 13 participating laboratories was confidentially assigned a participant number and image sets could be downloaded through a secure server. Results were distributed with each participant recognizable by its number and then each laboratory was able to modify their results with justification as model observer calculation are not yet a routine and potentially error prone.

Results: Detectability index increased with signal size for all participants and was very consistent for 6 mm sized target while showing higher variability for 8 and 10 mm sized target. There was one order of magnitude between the lowest and the largest uncertainty estimation.

Conclusions: This intercomparison helped define the state of the art of model observer performance computation and with thirteen participants, reflects openness and trust within the medical imaging community. The performance of a CHO with explicitly defined channels and a relatively large number of test images was consistently estimated by all participants. In contrast, the paper demonstrates that there is no agreement on estimating the variance of detectability in the training and testing setting.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9559858PMC
http://dx.doi.org/10.1002/mp.12940DOI Listing

Publication Analysis

Top Keywords

observer performance
12
model observer
12
participating laboratories
12
channelized hotelling
8
hotelling observer
8
model observers
8
performance computation
8
uncertainty estimation
8
sized target
8
image
6

Similar Publications

Background: In clinical practices, doctors usually need to synthesize several single-modality medical images for diagnosis, which is a time-consuming and costly process. With this background, multimodal medical image fusion (MMIF) techniques have emerged to synthesize medical images of different modalities, providing a comprehensive and objective interpretation of the lesion.

Purpose: Although existing MMIF approaches have shown promising results, they often overlook the importance of multiscale feature diversity and attention interaction, which are essential for superior visual outcomes.

View Article and Find Full Text PDF

Effective Nitrate Electroconversion to Ammonia Using an Entangled CoO/Graphene Nanoribbon Catalyst.

ACS Appl Mater Interfaces

December 2024

Institute of Chemistry, Federal University of Mato Grosso do Sul, Avenida Senador Filinto Muller 1555, Campo Grande, Mato Grosso do Sul 79074-460, Brazil.

There has been huge interest among chemical scientists in the electrochemical reduction of nitrate (NO) to ammonia (NH) due to the useful application of NH in nitrogen fertilizers and fuel. To conduct such a complex reduction reaction, which involves eight electrons and eight protons, one needs to develop high-performance (and stable) electrocatalysts that favor the formation of reaction intermediates that are selective toward ammonia production. In the present study, we developed and applied CoO/graphene nanoribbon (GNR) electrocatalysts with excellent properties for the effective reduction of NO to NH, where NH yield rate of 42.

View Article and Find Full Text PDF

Objective: The extent of resection (EOR) is an important prognostic factor for both low- and high-grade gliomas. Intraoperative MRI (iMRI) has been used to increase the EOR in glioma surgery. While a recent study reported differences between iMRI and early postoperative MRI (epMRI), their specific relationship to postoperative clinical symptoms remains unclear.

View Article and Find Full Text PDF

Hypertrophic cardiomyopathy (HCM) is a genetic cardiac disorder characterized by structural and functional abnormalities. Current management strategies, such as medications and septal reduction therapies, have significant limitations and risks. Recently, cardiac myosin inhibitors (CMIs) like mavacamten and aficamten have shown promise as noninvasive treatment options.

View Article and Find Full Text PDF

Purpose: To describe the clinical characteristics of macular holes (MH) with an intact bacillary layer.

Methods: This retrospective study analyzed the characteristics and surgical results of 23 eyes with MH and an intact bacillary layer, with or without rhegmatogenous retinal detachment (RRD).

Results: Among the 23 eyes, 11 had only MH, and 12 had concurrent fovea-off RRD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!