Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 143
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 143
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 209
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3051
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Severity: Warning
Message: Attempt to read property "Count" on bool
Filename: helpers/my_audit_helper.php
Line Number: 3053
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3053
Function: _error_handler
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: The effect of CSF on blood coagulation is not known. Enhanced coagulation by CSF may be an issue in thrombotic complications of ventriculoatrial and ventriculosinus shunts. This study aimed to assess the effect of CSF on coagulation and its potential effect on thrombotic events affecting ventriculovenous shunts.
Methods: Two complementary experiments were performed. In a static experiment, the effect on coagulation of different CSF mixtures was evaluated using a viscoelastic coagulation monitor. A dynamic experiment confirmed the amount of clot formation on the shunt surface in a roller pump model.
Results: CSF concentrations of 9% and higher significantly decreased the activated clotting time (ACT; 164.9 seconds at 0% CSF, 155.6 seconds at 9% CSF, and 145.1 seconds at 32% CSF). Increased clot rates (CRs) were observed starting at a concentration of 5% (29.3 U/min at 0% CSF, 31.6 U/min at 5% CSF, and 35.3 U/min at 32% CSF). The roller pump model showed a significantly greater percentage of shunt surface covered with deposits when the shunts were infused with CSF rather than Ringer's lactate solution (90% vs 63%). The amount of clot formation at the side facing the blood flow (impact side) tended to be lower than that at the side facing away from the blood flow (wake side; 71% vs 86%).
Conclusions: Addition of CSF to blood accelerates coagulation. The CSF-blood-foreign material interaction promotes clot formation, which might result in thrombotic shunt complications. Further development of the ventriculovenous shunt technique should focus on preventing CSF-blood-foreign material interaction and stagnation of CSF in wake zones.