Twiddler's Syndrome: A Rare Hardware Complication in Spinal Cord Stimulation.

Asian J Neurosurg

Department of Neurosurgery, Chosun University Hospital, College of Medicine, Chosun University, Gwangju, Republic of Korea.

Published: January 2018

Twiddler's syndrome is an uncommon hardware complication involving the lead and pulse generators in cardiac pacemakers and defibrillators, deep brain stimulators, and vagal nerve stimulators. However, until very recently, it had not been reported in spinal cord stimulation (SCS). Considering the incidence of hardware complications of spinal cord stimulation, there may be an underreporting of Twiddler's syndrome due to lack of awareness. Two cases of Twiddler's syndrome as a hardware complication of SCS were identified between 2005 and 2015. One patient with hardware failure due to Twiddler's syndrome refused to have a revision surgery. The other patient who had a lead migration associated with coiling of the lead and twisting of pulse generator needed a revision surgery. Twiddler's syndrome in patients treated with SCS is an uncommon but important adverse event. Awareness of characteristic presentation and radiologic finding is essential in the identification of Twiddler's syndrome in SCS.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5898115PMC
http://dx.doi.org/10.4103/ajns.AJNS_147_16DOI Listing

Publication Analysis

Top Keywords

twiddler's syndrome
28
hardware complication
12
spinal cord
12
cord stimulation
12
revision surgery
8
twiddler's
7
syndrome
6
hardware
5
syndrome rare
4
rare hardware
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!