Background: Corneal pachymetry is widely used for refractive surgery and follow up in keratoconus, accurate measurement is essential for a safe surgery.
Objective: To assess intraobserver reliability of central corneal thickness (CCT) measurements using optical low-coherence reflectometry (OLCR) technology and its agreement with ultrasonic pachymeter (US).
Method: Randomized and prospective comparative evaluation of diagnostic technology. One randomly healthy eye of subjects was scanned three times with both devices. Intraobserver within-subject standard deviation (Sw), coefficient of variation (CVw) and intraclass correlation coefficient (ICC) were obtained for reliability analysis; for study agreement, data were analyzed using the paired-sample t test and the Bland-Altman LoA method. The mean of three scans of each equipment was used to assess the LoA.
Results: The study enrolled 30 eyes of 30 subjects with average age of 28.70 ± 8.06 years. For repeatability, the Sw were 3.41 and 5.96 µ, the intraobserver CVw was 2 and 4% and ICC 0.991 and 0.988, for OLCR and US respectively. The mean CCT difference between OLCR and US was 8.90 ± 9.03 µ (95% confidence interval: 5.52-2.27 µ), and the LoA was 35.40 µ.
Conclusions: OLCR technology provided reliable intraobserver CCT measurements. Both pachymetry measurements may be used interchangeably with minimum calibration adjustment.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.24875/CIRU.M18000002 | DOI Listing |
Acta Parasitol
January 2025
Department of Molecular Biology and Genetics, Ordu University, Ordu, Turkey.
Purpose: Acanthamoeba species are eucaryotic protozoa found predominantly in soil and water. They cause ulceration and vision loss in the cornea (Acanthamoeba keratitis) and central nervous system (CNS) infection involving the lungs (granulomatous amoebic encephalitis). Antiparasitic drugs currently used in the treatment of infections caused by Acanthamoeba species are not effective at the desired level in some anatomical regions such as the eye and CNS.
View Article and Find Full Text PDFOphthalmol Glaucoma
January 2025
Glaucoma Division, Stein Eye Institute, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California. Electronic address:
Purpose: Investigate the influence of baseline blood pressure (BP) on retinal nerve fiber layer (RNFL) rates of change (RoC) in glaucoma patients with central damage or moderate to severe disease.
Design: Prospective cohort study.
Participants: 110 eyes with ≥4 RNFL optical coherence tomography scans and ≥2 years of follow-up.
Curr Eye Res
January 2025
Department of Ophthalmology, Harvey and Bernice Jones Eye Institute, University of Arkansas for Medical Sciences, Little Rock, AR, USA.
Purpose: To evaluate the effectiveness and safety of topical netarsudil 0.02% in managing childhood glaucoma.
Methods: A literature search in the electronic databases of PubMed CENTRAL, Google Scholar, EMBASE, the Register of Controlled Trials, and Ovid MEDLINE from January 2017 to August 2023 using one or a combination of the following terms: "netarsudil," "rhopressa," "Rho-kinase," "pediatric glaucoma," "childhood glaucoma," "intraocular pressure" was conducted.
J AAPOS
December 2024
Department of Ophthalmology, Boston Children's Hospital, Massachusetts Eye and Ear, Harvard Medical School, Boston, Massachusetts; Picower Institute for Learning and Memory, Massachusetts Institute of Technology, Cambridge, Massachusetts. Electronic address:
Ocular torticollis is traditionally attributed to eye misalignment, nystagmus, ptosis, or refractive error. We present 3 pediatric cases of acquired torticollis caused by a foreign body beneath the upper eyelid. The head posturing presumably developed to minimize contact of the foreign body with the corneal surface and mitigate ocular discomfort.
View Article and Find Full Text PDFZhonghua Yan Ke Za Zhi
January 2025
Department of Ophthalmology,Beijing Hospital, National Center of Gerontology Institute of Geriatric Medicine,Chinese Academy of Medical Sciences,Beijing100730,China.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!