Reliability and validity of MicroScribe-3DXL system in comparison with radiographic cephalometric system: Angular measurements.

Int Orthod

Center for restorative dentistry, Faculty of Dentistry, Universiti Teknologi MARA (UiTM), Jalan Hospital, Sungai Buloh, 47000 Selangor Darul Ehsan, Malaysia.

Published: June 2018

AI Article Synopsis

  • The study evaluated the reliability and validity of cephalometric measurements using the MicroScribe-3DXL with 60 Malay subjects, focusing on seven specific variables.
  • Significant differences were found in four of the measurements (U1/FH, FMA, IMPA, FMIA) with clinically acceptable variations, achieving high reliability (92.7%) and validity (91.8%) for the device.
  • The MicroScribe-3DXL is affirmed as a dependable tool in dental practice, offering time and cost savings, and shows potential for broader applications in research.

Article Abstract

Aim: The aim of the study was to assess the reliability and validity of cephalometric variables from MicroScribe-3DXL.

Materials And Methods: Seven cephalometric variables (facial angle, ANB, maxillary depth, U1/FH, FMA, IMPA, FMIA) were measured by a dentist in 60 Malay subjects (30 males and 30 females) with class I occlusion and balanced face. Two standard images were taken for each subject with conventional cephalometric radiography and MicroScribe-3DXL. All the images were traced and analysed. SPSS version 2.0 was used for statistical analysis with P-value was set at P<0.05.

Results: The results revealed a significant statistic difference in four measurements (U1/FH, FMA, IMPA, FMIA) with P-value range (0.00 to 0.03). The difference in the measurements was considered clinically acceptable. The overall reliability of MicroScribe-3DXL was 92.7% and its validity was 91.8%.

Conclusion: The MicroScribe-3DXL is reliable and valid to most of the cephalometric variables with the advantages of saving time and cost. This is a promising device to assist in diverse areas in dental practice and research.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ortho.2018.03.005DOI Listing

Publication Analysis

Top Keywords

reliability validity
8
cephalometric variables
8
validity microscribe-3dxl
4
microscribe-3dxl system
4
system comparison
4
comparison radiographic
4
cephalometric
4
radiographic cephalometric
4
cephalometric system
4
system angular
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!