The importance of correct implants positioning and masticatory load direction on a fixed prosthesis.

J Clin Exp Dent

DDs, MSc, PhD, Adjunct Professor, Department of Dental Materials and Proshodontics, São Paulo State University (Unesp), Institute of Science and Technology, São José dos Campos / SP, Brazil. Address: Av Engenheiro Francisco José Longo, 777, Jardim São Dimas, São José dos Campos, São Paulo, Brazil.

Published: January 2018

Background: Through the biomechanical study of dental implants, it is possible to understand the dissipation effects of masticatory loads in different situations and prevent the longevity of osseointegration. Aims: To evaluate the microstrains generated around external hexagon implants, using axial and non-axial loads in a fixed four-element prosthesis with straight implants and implants inclined at 17°.

Material And Methods: Three implants were modeled using CAD software following the manufacturer's measurements. Then, implants were duplicated and divided into two groups: one with straight implants and respective abutments, and the other with angled implants at 17° and respective abutments. Both groups were arranged inside a block simulating bone tissue. A simplified fixed prosthesis was installed on both groups and the geometries were exported to CAE software. Five loads of 300N were performed at axial and non-axial points on the fixed prosthesis. Stress on the implants and strain on the block were both analyzed. An in vitro experiment was performed following all structures made in FEA in order to validate the model. In each experimental block, 4 strain gauges were linearly placed between the implants and the same loads were repeated with a loading applicator device.

Results: The deformations computed by the gauges were correlated with the FEA results, showing that the group with inclined implants had more damaging biomechanical behavior and was significantly different from the group with straight implants (<0.005).

Conclusions: The mathematical model used is valid and inclined implants can induce unwanted bone remodeling. Finite Element Analysis, Dental Implants, Fixed Prosthesis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5899813PMC
http://dx.doi.org/10.4317/jced.54489DOI Listing

Publication Analysis

Top Keywords

fixed prosthesis
12
implants
12
straight implants
12
axial non-axial
8
respective abutments
8
correct implants
4
implants positioning
4
positioning masticatory
4
masticatory load
4
load direction
4

Similar Publications

Objectives: The objective of this study was to evaluate the long-term clinical outcomes of one-piece narrow-diameter implants (NDIs), with diameters of 2.5 mm and 3.0 mm, and to investigate the factors that affect marginal bone loss (MBL) around these implants.

View Article and Find Full Text PDF

Background: Acetabular chondrolabral delamination (ACD) is one of the most common hip cartilage injuries. However, there are very limited clinical treatments for this injury.

Purpose: To evaluate the effectiveness of poly (lactic-co-glycolic acid) (PLGA) conical nail fixation in the treatment of acute and chronic ACD in a porcine model.

View Article and Find Full Text PDF

Design, characterisation, and clinical evaluation of a novel porous Ti-6Al-4V hemipelvic prosthesis based on Voronoi diagram.

Biomater Transl

September 2024

Orthopaedic Research Institute and Department of Orthopaedics, West China Hospital, Sichuan University, Chengdu, Sichuan Province, China.

Three-dimensional printed Ti-6Al-4V hemipelvic prosthesis has become a current popular method for pelvic defect reconstruction. This paper presents a novel biomimetic hemipelvic prosthesis design that utilises patient-specific anatomical data in conjunction with the Voronoi diagram algorithm. Unlike traditional design methods that rely on fixed, homogeneous unit cell, the Voronoi diagram enables to create imitation of trabecular structure (ITS).

View Article and Find Full Text PDF

A technique to enhance the gingival esthetics of a prototype complete arch fixed implant-supported prototype prosthesis using a 3D printed gingival veneer.

J Prosthet Dent

December 2024

Clinical Professor and Program Director, University of Connecticut Health Center, Farmington, Conn.; and Private practice, Glastonbury, Conn. Electronic address:

An important step in the workflow for a complete arch fixed implant-supported prosthesis is the fabrication of a prototype resin prosthesis to allow the patient to visualize the definitive treatment outcome and also to use as an interim prosthesis. Producing a prototype prosthesis with the natural appearance of the gingival color and contour has been challenging. Current methods include painted stains, manually applied pink composite resin or autopolymerizing acrylic resin, and multicolored resin blocks, but these are all relatively expensive, have less than optimal esthetics, and are labor-intensive and time-consuming.

View Article and Find Full Text PDF

Background: Pelvic reconstruction after type I + II (or type I + II + III) internal hemipelvectomy with extensive ilium removal is a great challenge. In an attempt to anatomically reconstruct the hip rotation center (HRC) and achieve a low mechanical failure rate, a custom-made, 3D-printed prosthesis with a porous articular interface was developed. The aim of this study was to investigate the clinical outcomes of patients treated with this prosthesis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!