Background: Metal artifacts caused by high-density implants lead to incorrectly reconstructed Hounsfield units in computed tomography images. This can result in a loss of accuracy in dose calculation in radiation therapy. This study investigates the potential of the metal artifact reduction algorithms, Augmented Likelihood Image Reconstruction and linear interpolation, in improving dose calculation in the presence of metal artifacts.

Materials And Methods: In order to simulate a pelvis with a double-sided total endoprosthesis, a polymethylmethacrylate phantom was equipped with two steel bars. Artifacts were reduced by applying the Augmented Likelihood Image Reconstruction, a linear interpolation, and a manual correction approach. Using the treatment planning system Eclipse™, identical planning target volumes for an idealized prostate as well as structures for bladder and rectum were defined in corrected and noncorrected images. Volumetric modulated arc therapy plans have been created with double arc rotations with and without avoidance sectors that mask out the prosthesis. The irradiation plans were analyzed for variations in the dose distribution and their homogeneity. Dosimetric measurements were performed using isocentric positioned ionization chambers.

Results: Irradiation plans based on images containing artifacts lead to a dose error in the isocenter of up to 8.4%. Corrections with the Augmented Likelihood Image Reconstruction reduce this dose error to 2.7%, corrections with linear interpolation to 3.2%, and manual artifact correction to 4.1%. When applying artifact correction, the dose homogeneity was slightly improved for all investigated methods. Furthermore, the calculated mean doses are higher for rectum and bladder if avoidance sectors are applied.

Conclusion: Streaking artifacts cause an imprecise dose calculation within irradiation plans. Using a metal artifact correction algorithm, the planning accuracy can be significantly improved. Best results were accomplished using the Augmented Likelihood Image Reconstruction algorithm.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978555PMC
http://dx.doi.org/10.1002/acm2.12325DOI Listing

Publication Analysis

Top Keywords

augmented likelihood
20
likelihood image
20
image reconstruction
20
dose calculation
16
artifact correction
16
metal artifact
12
linear interpolation
12
irradiation plans
12
calculation radiation
8
radiation therapy
8

Similar Publications

Chronic coronary artery disease (CAD) remains a significant global healthcare burden. Current risk assessment methods have notable limitations in early detection and risk stratification. Hence, there is an urgent need for innovative biomarkers that facilitate the premature CAD diagnosis, ultimately leading to reduction in associated morbidity and mortality rates.

View Article and Find Full Text PDF

Recent Advances in Metal-Organic Framework-Based Anticancer Hydrogels.

Gels

January 2025

Exigo Recycling Pvt Ltd., Karnal 132114, Haryana, India.

Cancer is the second leading cause of death globally and the estimated number of new cancer cases and deaths will be ∼30.2 million and 16.3 million, respectively, by 2040.

View Article and Find Full Text PDF

Introduction: In clinical, the echocardiogram is the most widely used for diagnosing heart diseases. Different heart diseases are diagnosed based on different views of the echocardiogram images, so efficient echocardiogram view classification can help cardiologists diagnose heart disease rapidly. Echocardiogram view classification is mainly divided into supervised and semi-supervised methods.

View Article and Find Full Text PDF

Background: Despite the well-documented immune dysregulation in both psoriasis and Sjögren's syndrome (SS), the specific link between these two autoimmune diseases has not been extensively explored. The present study aims to investigate the impact of psoriasis on the risk of SS.

Methods: A retrospective cohort study using TriNetX data compared SS development in patients with psoriasis and controls using propensity score matching, Kaplan-Meier curves, and Cox models.

View Article and Find Full Text PDF

Purpose: Head acceleration events (HAEs) are a growing concern in contact sports, prompting two rugby governing bodies to mandate instrumented mouthguards (iMGs). This has resulted in an influx of data imposing financial and time constraints. This study presents two computational methods that leverage a dataset of video-coded match events: cross-correlation synchronisation aligns iMG data to a video recording, by providing playback timestamps for each HAE, enabling analysts to locate them in video footage; and post-synchronisation event matching identifies the coded match event (e.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!