AI Article Synopsis

  • Stenting is preferred for treating obstructive venous disease over balloon angioplasty due to better outcomes, prompting a study on a new venous scaffold called Petalo CVS.
  • In the study, 12 healthy pigs were used to evaluate the safety and efficacy of the Petalo CVS, with successful implantation and no complications noted during follow-ups.
  • Results showed a 100% primary patency rate and minimal inflammatory response, indicating that Petalo CVS may be a safe and effective option for venous treatment.

Article Abstract

Background: Stenting has become the first-line treatment of obstructive venous disease because of poor results of balloon angioplasty. This preclinical study aimed to investigate the safety and efficacy profile of a novel compliant venous scaffold (CVS) denominated Petalo CVS, specifically designed for venous diseases.

Materials And Methods: Twelve healthy pigs weighing 90 kg were used to test Petalo CVS. The devices were implanted into the internal jugular veins (IJVs) using a femoral vein percutaneous approach. The safety profile including the success rate of device releasing, anchoring, and positioning was evaluated immediately. Fracture, migration, primary patency, and endothelial response were assessed at 1, 2, 3, and 6 months after the study procedure.

Results: A total of 32 devices were successfully released in both IJVs. No procedure- or device-related complications were reported, and all pigs successfully completed the different scheduled follow-up periods. The primary patency rate was 100%, and no fracture or migration of the device into the brachiocephalic trunk was reported. Histological examination revealed only minimal lesions with minimal or absent inflammatory reaction surrounding the incorporated metallic rods.

Conclusions: This porcine model study showed a promising safety and efficacy profile of Petalo CVS, a novel endovenous device based on specific concepts.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5832035PMC
http://dx.doi.org/10.1155/2018/7312315DOI Listing

Publication Analysis

Top Keywords

petalo cvs
12
novel compliant
8
porcine model
8
model study
8
safety efficacy
8
efficacy profile
8
fracture migration
8
primary patency
8
compliant scaffold
4
scaffold specific
4

Similar Publications

Article Synopsis
  • Stenting is preferred for treating obstructive venous disease over balloon angioplasty due to better outcomes, prompting a study on a new venous scaffold called Petalo CVS.
  • In the study, 12 healthy pigs were used to evaluate the safety and efficacy of the Petalo CVS, with successful implantation and no complications noted during follow-ups.
  • Results showed a 100% primary patency rate and minimal inflammatory response, indicating that Petalo CVS may be a safe and effective option for venous treatment.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!