Objective: This study evaluated the imaging characteristics and accuracy of T2-weighted (T2W) balanced steady-state free procession (b-SSFP) magnetic resonance imaging, relative to b-SSFP or single-shot fast spin echo (SSFSE), for the diagnosis of placental adhesion disorder (PAD).
Methods: Fifty-one pregnant patients suspected of PAD were examined with T2W b-SSFP, b-SSFP and SSFSE. The image types were independently analysed for signs of PAD: abnormal placental bulge (APB), dark intraplacental bands (DIB), placental heterogeneity (PH) and placental protrusion into adjacent structures (PPAS). The sequences were compared for muscle-to-placenta signal ratio, signs of PAD and area under the receiver operating characteristic curve (AUC) for diagnostic accuracy of PAD.
Results: PAD was confirmed in 34 women. The muscle-to-placenta signal ratio was highest in the T2W b-SSFP. The diagnostic rates of APB in T2W b-SSFP were comparable to that of b-SSFP, but were significantly higher than that of SSFSE. The rates of PH in SSFE were comparable to that of b-SSFP, but both were significantly lower than that of T2W b-SSFP. The rates of DIB were significantly higher in T2W b-SSFP images compared with SSFSE. Rates of PPAS were comparable among three sequences. The AUCs of the T2W b-SSFP, b-SSFP and SSFSE were 0.966, 0.890 and 0.823, respectively.
Conclusion: T2W b-SSFP has high diagnostic accuracy for PAD relative to SSFSE or b-SSFP, which may be due to its high SNR, T2-weighting and lack of blur.
Key Points: • Signal myometrium-to-placenta ratio was highest in the T2W b-SSFP images. • Diagnostic rate of APB in T2W b-SSFP was highest. • Diagnostic rate of DIB was higher in T2W b-SSFP than in SSFSE. • Diagnostic rate of PH in T2W b-SSFP was highest. • Maximum AUC for diagnostic accuracy of PAD was in T2W b-SSFP.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00330-018-5388-0 | DOI Listing |
Eur Radiol
September 2018
MR department of Affiliated Zhongshan City Hospital of Sun Yat-sen University, Sun Wendong Road No. 2, Zhongshan City, Guangdong Province, China.
Objective: This study evaluated the imaging characteristics and accuracy of T2-weighted (T2W) balanced steady-state free procession (b-SSFP) magnetic resonance imaging, relative to b-SSFP or single-shot fast spin echo (SSFSE), for the diagnosis of placental adhesion disorder (PAD).
Methods: Fifty-one pregnant patients suspected of PAD were examined with T2W b-SSFP, b-SSFP and SSFSE. The image types were independently analysed for signs of PAD: abnormal placental bulge (APB), dark intraplacental bands (DIB), placental heterogeneity (PH) and placental protrusion into adjacent structures (PPAS).
Eur J Radiol
May 2010
Department of Neuroradiology, Geneva University Hospital and University of Geneva, Geneva, Switzerland.
Imaging plays an essential role for the detection and analysis of pathologic conditions of the brachial plexus. Currently, several new techniques are used in addition to conventional 2D MR sequences to study the brachial plexus: the 3D STIR SPACE sequence, 3D heavily T2w MR myelography sequences (balanced SSFP=CISS 3D, True FISP 3D, bFFE and FIESTA), and the diffusion-weighted (DW) neurography sequence with fiber tracking reconstruction (tractography). The 3D STIR sequence offers complete anatomical coverage of the brachial plexus and the ability to slice through the volume helps to analyze fiber course modification and structure alteration.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!