Background: One of the most critical points in the planning of hip replacement surgeries is the selection of the implant, since its survival significantly impacts the patients health. However, the ideal survival time that an implant must prove to be selected has not been defined. The objective of this consensus is to define minimum performance standards for the selection of conventional primary hip replacement prosthesis.

Material And Methods: The consensus was carried out using the methodology of nominal group. This included: 1. A review of the available evidence and the issues to be evaluated, 2. Meeting for the vote and discussion, 3. Quantitative statistical analysis with median (M) and interquartile range (IQR) and qualitative one with proportions of the results to generate recommendations.

Results: The primary source of information for prosthesis selection (M: 8; IQR: 7-9), choice in the event of conflicting evidence (M: 8; IQR: 7-9), or limited evidence in the literature (M: 7; IQR: 4.75-825) should be national registries. The minimum acceptable follow-up is 10 years (M: 9; IQR: 8-9) and the minimum acceptable survival is 90% at 10 years (M: 8; IQR: 5-8.5).

Discussion: According to these results, the consensus of experts proposed that the selection of the implant for conventional primary hip replacement must be based on the information published in the national registries and that the prosthesis must have a minimum follow-up of 10 years and show a minimum survival of 90%.

Download full-text PDF

Source

Publication Analysis

Top Keywords

primary hip
12
minimum acceptable
12
hip replacement
12
acceptable survival
8
selection implant
8
conventional primary
8
iqr 7-9
8
national registries
8
follow-up years
8
years iqr
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!