Atrial fibrillation (AF) ablation requires transseptal puncture to access the left atrium. Recently, a radiofrequency (RF) needle was developed. The purpose of this study was to compare the incidence of MRI-confirmed acute cerebral embolism (ACE) during AF ablation procedures performed with RF needle versus mechanical needle transseptal puncture. This study consisted of 383 consecutive patients who underwent catheter ablation for AF that required transseptal puncture with mechanical or radiofrequency transseptal needles. Of those, 232 propensity score-matched patients (116 with each needle type) were included in the analysis. All patients had cerebral MRI performed 1 or 2 days after the procedure. Baseline characteristics were similar between the two groups. Total procedure time was significantly shorter in Group RF than Group non-RF (167 ± 50 vs. 181 ± 52 min, P = 0.01). ACE was detected by MRI in 59 (25%) patients. All patients with ACE were asymptomatic. Incidence of ACE was lower in Group RF than Group non-RF (19 vs. 32%, P = 0.02). B-type natriuretic peptide level was higher in the patients with ACE as compared to those without ACE (65.2 ± 68.7 vs. 44.7 ± 55.1 pg/ml, P = 0.02). In multivariable analysis, the use of RF needle and BNP level was related to the incidence of ACE (OR = 0.499, 95% CI 0.270-0.922, P = 0.03 and OR = 1.005, 95% CI 1.000-1.010, P = 0.03). Use of RF needle for transseptal puncture was associated with lower total procedure time and risk of ACE during catheter ablation of AF.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00380-018-1159-8DOI Listing

Publication Analysis

Top Keywords

transseptal puncture
20
needle transseptal
12
catheter ablation
12
radiofrequency needle
8
puncture associated
8
associated lower
8
atrial fibrillation
8
propensity score-matched
8
ace
8
total procedure
8

Similar Publications

Background: In patients with mechanical aortic and mitral valves requiring catheter ablation of ventricular tachycardia (VT), a technique for access from the right atrium (RA) to the left ventricle (LV) via puncture of the inferoseptal process of the LV was previously described in a single-center series.

Objectives: This study sought to report the multicenter experience of VT ablation using this novel LV access approach.

Methods: We assembled a multicenter registry of patients with double mechanical valves who underwent VT ablation with RA-to-LV access.

View Article and Find Full Text PDF

This case report discusses the management of a 75-year-old man who developed an unusual type of atypical atrial flutter following a previous pulmonary vein isolation for paroxysmal atrial fibrillation. Despite a second attempt to re-isolate the pulmonary veins and performing cavotricuspid isthmus ablation (which was suspected to be part of the arrythmia circuit), the flutter continued and was converted to sinus rhythm through electrical cardioversion. A few weeks later, the patient's atrial tachycardia relapsed.

View Article and Find Full Text PDF

: The safety and efficacy of electrophysiological (EP) procedures using ultrasound (US) guidance are being increasingly studied. We investigated if a systematic workflow with ultrasound guidance (the US4ABL), comprising four steps (transesophageal echocardiography (TEE) for left atrial thrombus exclusion, US of the groin vessels to guide femoral access, TEE-aided transseptal puncture, and transthoracic echocardiography (TTE) for exclusion of pericardial tamponade after the procedure), reduces the number of complications and fluoroscopy duration and dose. : A total of 212 consecutive patients underwent left-sided ablations using the US4ABL workflow and were compared to a group of 299 patients who underwent the same type of ablations using post-procedural TTE to exclude tamponade (standard group: venous and/or arterial access by palpation and fluoroscopy, and pressure guided transseptal puncture).

View Article and Find Full Text PDF

Transjugular transcatheter edge-to-edge mitral valve repair in a patient with functional mitral regurgitation: a case report.

Eur Heart J Case Rep

January 2025

Department of Cardiovascular Medicine, Sendai Kousei Hospital, 1-20 Tsutsumidori-amamiya, Aoba Ward, Sendai, Miyagi 9810914, Japan.

Article Synopsis
  • The transcatheter edge-to-edge mitral valve repair (M-TEER) using the MitraClip system is usually done via the transfemoral approach, but the transjugular approach can be a viable alternative when the former isn't possible.
  • A 57-year-old man with heart failure and severe functional mitral regurgitation could not undergo the transfemoral approach due to an occluded inferior vena cava, leading to the decision to use the transjugular technique instead.
  • The procedure was successful with no complications, significantly reducing the patient's mitral regurgitation, and he showed sustained improvement a year later.
View Article and Find Full Text PDF

Intracardiac echocardiography (ICE) has been used to guide radio-frequency catheter ablation (RFCA) for better catheter navigation and less radiation exposure in treating atrial fibrillation (AF). This retrospective cohort study enrolled 227 AF patients undergoing ICE- or traditional fluoroscopy (TF)-guided RFCA for AF in a tertiary hospital. ICE was used more often in patients with atrial tachycardia [odds ratio (OR) 3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!