Aim: We compared early postprocedural and midterm evolution of atrioventricular and intraventricular conduction disorders following implantation of the new generation Evolut R (ER) prosthesis in comparison with the previous generation CoreValve (CV) system using routinely recorded ECG up to 6-month follow-up.

Methods: All consecutive patients treated by transcathether aortic valve implantation (TAVI) using the Medtronic self-expanding devices for symptomatic severe aortic stenosis in a single centre between October 2011 and February 2016 were considered for inclusion. ECGs recorded at baseline, day 1 after TAVI, discharge and 6 months were retrospectively analysed. At each time-point, intrinsic rhythm, PR interval, QRS axis and duration, and atrioventricular and intraventricular conduction were analysed. Atrioventricular and intraventricular conduction following TAVI at discharge and at 6 months were compared intrasubject at the different time intervals and between patients receiving the ER versus the CV prosthesis.

Results: Among the 113 patients included in the analysis (51% female, 83.3±6.2 years), 60 (53%) patients received the CV and 53 (47%) patients received the ER. Compared with patients in the CV group, those in the ER group had a lower Society of Thoracic Surgeons score (6.3±3.1vs 4.8±3.6, P=0.02). Patients in the ER group in comparison with those in the CV group more frequently had postprocedural PR interval (57%vs23%, respectively, P=0.004) and QRS prolongation (76%vs50%, P=0.03) at discharge. Incidence of complete atrioventricular block was similar between both groups (9%vs18%, P=0.3) up to 6-month follow-up. No difference in term of new left bundle branch block (LBBB) (34%vs28%, P=0.8) or permanent pacemaker implantation rates (32.1%vs31.7%, P=1.0) was reported.

Conclusions: Patients with the ER had greater postprocedural atrioventricular and intraventricular conduction delays than those with the CV at discharge, with however similar incidence of high-degree atrioventricular block, new LBBB and permanent pacemaker implantation up to 6-month follow-up.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5888433PMC
http://dx.doi.org/10.1136/openhrt-2017-000770DOI Listing

Publication Analysis

Top Keywords

atrioventricular intraventricular
16
intraventricular conduction
16
conduction disorders
8
evolut prosthesis
8
patients
8
tavi discharge
8
discharge months
8
patients received
8
patients group
8
discharge incidence
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!