Background: Carcinoembryonic antigen (CEA) is one of the tumor markers available for evaluating disease progression status after initial therapy and monitoring subsequent treatment modalities in colorectal, gastrointestinal, lung, and breast carcinoma. We evaluated the correlations and differences between widely used, automated CEA immunoassays at four different medical laboratories.
Methods: In total, 393 serum samples with CEA ranging from 3.0 to 1,000 ng/mL were analyzed on ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY, USA), ARCHITECT i2000sr (Abbott Diagnostics, Abbott Park, IL, USA), Elecsys E170 (Roche Diagnostics, Indianapolis, IN, USA), and Unicel DxI800 (Beckman Coulter, Fullerton, CA, USA), and the results were compared. Deming regression, Passing-Bablok regression, and Bland-Altman analyses were performed to evaluate the data correlation and % differences among these assays.
Results: Deming regression analysis of data from Elecsys E170 and UniCel DxI800 showed good correlation (y=3.1615+0.8970x). According to Bland-Altman plot, no statistically significant bias (-1.78 ng/mL [95% confidence interval: -4.02 to 0.46]) was observed between Elecsys E170 and UniCel DxI800. However, the relative differences of CEA concentrations between assays exceeded the acceptable limit of 30%. Regarding the agreement of positivity with cut-off value 5.0 ng/mL, ARCHITECT i2000sr and Elecsys E170 showed the highest agreement (95.2%), whereas ADVIA Centaur XP and ARCHITECT i2000sr showed the lowest agreement (70.7%).
Conclusions: Agreements between automated CEA immunoassays are variable, and individual CEA concentrations may differ significantly between assays. Standardization of serum CEA concentrations and further harmonization are needed.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5895865 | PMC |
http://dx.doi.org/10.3343/alm.2018.38.4.355 | DOI Listing |
Front Immunol
August 2022
Department of Clinical Laboratory, The First People's Hospital of Wenling, Affiliated to Wenzhou Medical University, Taizhou, China.
Objective: To evaluate a novel fully automated immunoturbidimetric assay developed by Qiangsheng Biotechnology Company for the detection of anticyclic citrullinated peptide antibodies (anti-CCP) in serum of patients with rheumatoid arthritis (RA) and compare it to the conventional EUROIMMUN- anti-CCP ELISA. Two other commonly used automated assays, the Elecsys anti-CCP assay, an ECLIA that is run on the Modular Analystics E170 (Cobas Diagnostics, Germany), and an anti-CCP CLIA developed by YHLO that is run on the iFlash 3000 Chemiluminescence Immunoassay Analyzer, were included as reference standards.
Methods: A total of 264 serum samples were collected from patients attending the First People's Hospital of Wenling affiliated to Wenzhou Medical University between July 2020 and November 2020.
Ann Lab Med
July 2018
Department of Laboratory Medicine, College of Medicine, The Catholic University of Korea, Seoul, Korea.
Background: Carcinoembryonic antigen (CEA) is one of the tumor markers available for evaluating disease progression status after initial therapy and monitoring subsequent treatment modalities in colorectal, gastrointestinal, lung, and breast carcinoma. We evaluated the correlations and differences between widely used, automated CEA immunoassays at four different medical laboratories.
Methods: In total, 393 serum samples with CEA ranging from 3.
J Clin Lab Anal
January 2016
Service de Biochimie métabolique, Groupe hospitalier Pitié Salpêtrière, 47-83 bd de l'Hôpital, 75013, Paris, France.
Background: Tumor marker measurements are becoming essential for prognosis and follow-up of patients in oncology. In this context, we aimed to compare a new analyzer, Lumipulse(®) G1200 (Fujirebio group, distributed in Europe by the Innogenetics group) with Kryptor(®) (Thermo Fisher Scientific B.R.
View Article and Find Full Text PDFAnn Clin Biochem
May 2015
General Laboratory, Regional University Hospital, Málaga, Spain.
Background: Immunometric assays have recently shown higher specificity in the detection of human insulin than radioimmunoassays with almost no cross-reaction with proinsulin or C peptide. The introduction of the new insulin analogues on the market, however, has raised the need to define their cross-reactivity in these assays. Several studies have been published in this regard with different results.
View Article and Find Full Text PDFJ Virol Methods
April 2014
Fondazione IRCCS, Policlinico San Matteo, Struttura Complessa di Ostetricia e Ginecologia, Pavia, Italy.
Screening for acute rubella infection in pregnancy is an important element of antenatal care. This study compared the sensitivity, specificity and reproducibility of two new, fully automated Elecsys(®) Rubella IgM and IgG immunoassays designed for the Elecsys 2010, Modular Analytics E170, COBAS e-411 and COBAS e-601 and e602 analytical platforms, with current assays using serum from patients with primary rubella infections, vaccinated patients, patients with potentially cross-reacting infections and on routine samples in clinical laboratories in France, Germany and Italy. Both assays showed good within-run and within-laboratory precision.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!