Background: Observational studies have explored the safety of magnetic resonance (MR) scanning of patients with cardiac implantable electronic devices (CIEDs) that are not Food and Drug Administration approved for MR scanning ("nonconditional"). However, concern has been raised that MR scanning that includes the thoracic region may pose a higher risk. This study examines the safety of MR scanning of thoracic versus nonthoracic regions of patients with CIEDs.

Methods: Patients underwent MR scanning utilizing an institutional protocol. CIED variables examined included sensing value, pacing capture threshold, lead impedance, and battery voltage. Regression analysis of the CIED variable differences (pre- to immediately post-MR and pre-MR to long-term follow-up) was performed to determine if CIED variable differences were dependent on region scanned (thoracic vs nonthoracic), time from CIED implant to MR scanning, or CIED type (pacemaker vs implantable cardioverter defibrillator).

Results: 238 patients (38% female, age 65 ± 15 years) underwent 339 MR scans, including 99 MR scans of the thoracic region. CIED variable differences to immediately post-MR or to long-term follow-up were not significantly different from zero (P > 0.05) and there was no dependence upon region scanned (thoracic vs nonthoracic), time from CIED implant to MR scan, or CIED type. One power-on reset occurred in a patient that underwent a cardiac MR and the CIED was successfully reprogrammed. There were no clinical adverse effects.

Conclusions: CIED variable differences following MR scan were not dependent on the region scanned (thoracic vs nonthoracic) and there were no clinical adverse effects in this prospective cohort.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.13340DOI Listing

Publication Analysis

Top Keywords

cied variable
16
variable differences
16
region scanned
12
scanned thoracic
12
thoracic nonthoracic
12
cied
10
thoracic versus
8
versus nonthoracic
8
cardiac implantable
8
implantable electronic
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!