Objective: The objective of this study was to report the 1-year follow-up study results of the new Horizon stent graft (Endospan, Herzliya, Israel) from two different prospective consecutive trials. The Horizon abdominal aortic aneurysm stent graft system is a 14F profile system requiring only a single access site. It consists of three modules, introduced separately: base limb (iliac to iliac limb); distal aortic limb; and proximal aortic limb with a bare suprarenal crown and active fixation.
Methods: Data from the first in man (FIM) clinical study with 10 patients enrolled and the pivotal study with 30 patients were analyzed. Outcomes measured were freedom from major adverse events (MAEs) including all-cause mortality, myocardial infarction, renal failure, respiratory failure, paraplegia, stroke, bowel ischemia, and procedural blood loss ≥1000 mL. Performance end points included successful delivery and deployment of the device, freedom from aneurysm growth ≥5 mm, type I or type III endoleak, stent graft occlusion, conversion to open surgery, rupture, and stent graft migration.
Results: In the FIM study, one conversion to open surgery with >1000 mL of blood loss was registered perioperatively. In the pivotal study, no perioperative MAE was registered. Overall, at 1-year follow-up, two deaths and one aneurysm growth unrelated to endoleak were registered.
Conclusions: The results of both the FIM and pivotal studies demonstrated that 39 of 40 procedures were successful for delivery and deployment of the Horizon stent graft. No MAE was registered during the follow-up. The primary safety and performance end points were met in both studies.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jvs.2017.11.069 | DOI Listing |
J Cardiovasc Surg (Torino)
February 2025
Department of Vascular Surgery, AZ Sint Blasius, Dendermonde, Belgium.
Background: Due to the heterogeneity of literature findings, stent type selection for the endovascular treatment of complex aorto-iliac occlusive disease remains challenging. The BELSTREAM study, a physician-initiated, prospective, multicenter, single-arm study, aims to report the safety and efficacy of the balloon expandable LifeStream Peripheral Stent Graft System (BD, Tempe, Arizona, USA) for the treatment of complex TASC C and D aorto-iliac artery lesions.
Methods: Seventy patients and 133 lesions were included at six Belgian institutions.
Ann Vasc Dis
January 2025
Department of Vascular Surgery, The University of Tokyo, Tokyo, Japan.
The underlying mechanisms of abdominal aortic aneurysms (AAAs) are not fully understood. Given the multifactorial nature of AAA development and progression, a comprehensive approach is essential. Throughout my academic career, I conducted various studies on AAA.
View Article and Find Full Text PDFJ Cardiothorac Surg
January 2025
Prague Lung Transplant Program, 3rd Department of Surgery, First Faculty of Medicine, Charles University and Motol University Hospital, Prague, Czech Republic.
Scoliotic deformity represents a serious spinal disorder that influences the locomotive and cardiopulmonary systems. Some patients with severe scoliosis and end-stage lung disease are therefore denied lung transplantation. In patients with scoliosis considering lung transplantation, size match, straight back syndrome, delayed chest closure and bronchial stenosis are key issues clinicians should evaluate.
View Article and Find Full Text PDFCureus
December 2024
Medicine/Cardiology, Rutgers University New Jersey Medical School, Newark, USA.
Coronary artery disease (CAD) is associated with poor outcomes after orthotopic liver transplantation (OLT). We report on six high-risk end-stage liver disease (ESLD) patients who underwent percutaneous coronary intervention (PCI) with bare metal stents during the preoperative evaluation process. There was no mortality or major adverse cardiac event (MACE) within 90 days of OLT.
View Article and Find Full Text PDFTrauma Case Rep
February 2025
Department of Emergency and Critical Care Medicine, Hyogo Emergency Medical Center, 1-3-1 Wakinohamakaigandori, Chuo-ku, Kobe, Hyogo 651-0073, Japan.
Background: Hybrid emergency rooms (ERs) allow computed tomography (CT) scanning, interventional radiology, and surgery all in the same suite. Severe trauma patients with blunt thoracic aortic injury (BTAI) require rapid diagnosis and treatment. Hybrid ERs allow the potential for clinicians to implement multiple therapeutic procedures, including thoracic endovascular aortic repair (TEVAR), for these types of conditions without the need to transport the patients.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!