Introduction: Measurement of physiological parameters in extreme environments is essential to advancing knowledge, prophylaxis and treatment of altitude sickness. Point-of-care testing facilitates investigation in non-specialist and remote settings, as well as becoming increasingly popular at the bedside for real-time results in the clinical environment. Arterialised capillary earlobe blood gases are recommended as a valid alternative to arterial sampling in research. This study aimed to test the feasibility of obtaining and analysing daily earlobe samples at high altitude.

Methods: From 17 to 24 January 2016, 24 participants on a research expedition to Ecuador underwent daily earlobe blood gas measurements including pH, partial pressure of oxygen and partial pressure of carbon dioxide to 5043 m. Samples were analysed using an Abbott i-STAT blood gas analyser and G3+ cartridges.

Results: Daily measurements were successfully obtained and analysed at the point of care in 23/24 participants and were well tolerated with no adverse events. 12% (27/220) cartridges failed and required repeat sampling.

Conclusions: Daily earlobe blood gas analysis using the Abbott i-STAT is feasible in a protected environment at high altitude. Participants and equipment should be kept warm before and during testing. Spare cartridges should be available. This methodology may be useful for both research and therapeutic measurements in remote, rural and wilderness medicine.

Download full-text PDF

Source
http://dx.doi.org/10.1136/jramc-2017-000902DOI Listing

Publication Analysis

Top Keywords

earlobe blood
16
blood gas
16
abbott i-stat
12
daily earlobe
12
high altitude
8
arterialised capillary
8
capillary earlobe
8
partial pressure
8
earlobe
5
blood
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!