Quality assurance in CT: implementation of the updated national diagnostic reference levels using an automated CT dose monitoring system.

Clin Radiol

Department of Diagnostic and Interventional Radiology, Medical Faculty, University Dusseldorf, Moorenstraße 5, 40225 Dusseldorf, Germany.

Published: July 2018

Aim: To evaluate the implementation of the updated computed tomography (CT) diagnostic reference levels (DRLs) from the German Federal Office for Radiation Protection into clinical routine using an automatic CT dose monitoring system.

Methods And Materials: CT radiation exposure was analysed before and after implementing the updated national DRLs into routine clinical work in 2016. After the implementation process, institutional CT protocols were mapped to the anatomical regions for which DRLs were provided. Systematically, protocols that exceeded the thresholds were optimised and analysed in detail. The CT radiation output parameters analysed were volumetric CT dose index (CTDIvol) and dose-length product (DLP). Three radiologists evaluated subjective image quality using a three-point Likert scale.

Results: The study included 94,258 CT series (from 27,103 CT examinations) in adult patients performed in 2016. When averaged over all body regions with available DRL, institutional CTDIvol/DLP values were always below the DRLs (65.2±32.9%/67.3±41.5% initially; 59.4±32%/60.5±39.9% after optimisation). Values exceeding the national DRLs were found for pelvis (n=268; CTDIvol 107.7±65.7%/DLP 106.3±79.3%), lumbar spine (n=91; 160.8±74.7%/175.2±104.1%), and facial bones (n=527; 108±39%/152.7±75.7%). After optimisation, CTDIvol and DLP were 87.9±73%/87.8±80.8% for the pelvis, 67.8±33.2%/74.5±50.6% for the lumbar spine and 95.1±45.8%/133.3±74.6% for the viscerocranium.

Conclusion: An automatic CT dose monitoring system enabled not only comprehensive monitoring of a DRL implementation process but can also help to optimise radiation exposure.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.crad.2018.02.012DOI Listing

Publication Analysis

Top Keywords

dose monitoring
12
implementation updated
8
updated national
8
diagnostic reference
8
reference levels
8
monitoring system
8
automatic dose
8
radiation exposure
8
national drls
8
implementation process
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!