AI Article Synopsis

  • Flexible ureterorenoscopy (fURS) is effective for treating kidney stones under 2 cm and upper urinary tract cancers, but there's limited research on which fURS provides the best image quality.
  • A study compared seven types of fURS using videos rated by 103 participants, revealing that digital ureterorenoscopes generally outperformed fiber optic ones in image quality.
  • The Storz D ureterorenoscopes, especially in Clara Chroma mode, provided the best overall image quality, with Olympus V performing notably well, while Olympus P6 and Storz X2 had lower quality ratings.

Article Abstract

Introduction: Flexible ureterorenoscopy (fURS) is one of the best solutions for treatment of renal calculi <2 cm and for upper urinary tract urothelial carcinoma conservative treatment. An adequate quality of vision is mandatory to help surgeon get better outcomes. No studies have been done, to our knowledge, about what fURS in the market has the best quality of vision.

Materials And Methods: Seven different fURS were used to compare the image quality (Lithovue, Olympus V, Olympus V2, Storz Flex XC-in White Light and in Clara+Chroma mode-Wolf Cobra Vision, Olympus P6, and Storx Flex X2). Two standardized grids to evaluate contrast and image definition and three stones of different composition were filmed in four standardized different scenarios. These videos were shown to 103 subjects (51 urologists and 52 nonurologists) who had to evaluate them with a rating scale from 1 (very bad) to 5 (very good).

Results: No difference in terms of scores was observed for sex of the participants. Digital (D) ureterorenoscopes were rated better than fiber optics (FOs) ureterorenoscopes. Overall, Flex XC White Light and XC Clara+Chroma image quality resulted steadily better than other fURS (p < 0.0001). Olympus V generally provided a vision better than Lithovue. Cobra Vision and Olympus V2 had superimposable values that were significantly lower than Lithovue's ones. Olympus P6 and Storz X2 offered a low quality of vision compared to the others. In the medium simulating bleeding, Olympus V and V2 significantly improved their scores of 12% and 8.1%, contrary to rest of the ureterorenoscopes.

Conclusion: D ureterorenoscopes have a better image quality than FO ones. The only disposable ureterorenoscope tested was comparable to the majority of other D ureterorenoscopes. The best image quality was provided by Storz D ureterorenoscopes, being Clara Chroma the favorite Spies Mode, according to literature.

Download full-text PDF

Source
http://dx.doi.org/10.1089/end.2017.0838DOI Listing

Publication Analysis

Top Keywords

comparison flexible
4
flexible ureterorenoscope
4
ureterorenoscope quality
4
quality vision
4
vision vitro
4
vitro study
4
study introduction
4
introduction flexible
4
flexible ureterorenoscopy
4
ureterorenoscopy furs
4

Similar Publications

Inorganic pyrophosphatases, or PPases, are ubiquitous enzymes whose activity is necessary for a large number of biosynthetic reactions. The catalytic function of PPases is dependent on certain conformational changes that have been previously characterized based on the comparison of the crystal structures of various complexes. The current work describes the conformational dynamics of a structural model of human mitochondrial pyrophosphatase hPPA2 using molecular dynamics simulation, all-atom principal component analysis, and coarse-grained normal mode analysis.

View Article and Find Full Text PDF

This study investigates the mechanical properties of thermoplastic polyurethane (TPU) 60A, which is a flexible material that can be used to produce soft robotic grippers using additive manufacturing. Tensile tests were conducted under ISO 37 and ISO 527 standards to assess the effects of different printing orientations (0°, 45°, -45°, 90°, and quasi-isotropic) and test speeds (2 mm/min, 20 mm/min, and 200 mm/min) on the material's performance. While the printing orientations at 0° and quasi-isotropic provided similar performance, the quasi-isotropic orientation demonstrated the most balanced mechanical behavior, establishing it as the optimal choice for robust and predictable performance, particularly for computational simulations.

View Article and Find Full Text PDF

Sexual dimorphism of the humerus bones in a French sample: comparison of several statistical models including machine learning models.

Int J Legal Med

January 2025

Faculté de Médecine, Institut Universitaire d'Anthropologie Médico-Légale, Université Côte d'Azur, 28 Avenue de Valombrose, Nice Cedex 2, 06107, France.

Sex estimation is an important part of skeletal analysis and forensic identification. Traditionally pelvic traits are utilized for accurate sex estimation. However, the long bones, especially humerus, have been proved to be as effective for determine the sex of the individual.

View Article and Find Full Text PDF

We study the problem of testing multiple secondary endpoints conditional on a primary endpoint being significant in a two-stage group sequential procedure, focusing on two secondary endpoints. This extends our previous work with one secondary endpoint. The test for the secondary null hypotheses is a closed procedure.

View Article and Find Full Text PDF

With the goal of fast and accurate diagnosis of infectious diseases, this study presents a novel electrochemical biosensor that employs a refined aptamer (C9t) for the detection of spike (S) protein SARS-CoV-2 variants in a flexible multielectrode aptasensor array with PoC capabilities. Two aptamer modifications were employed: removing the primer binding sites and including two dithiol phosphoramidite anchor molecules. Thus, reducing fabrication time from 24 to 3 h and increasing the stability and sparseness for multi-thiol aptasensors compared to a standard aptasensor using single thiols, without a reduction in aptamer density.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!