Background: Despite the development of various methods of intraocular pressure (IOP) measurement, Goldmann applanation tonometry (GAT) is still the most popular. The measurement using GAT depends on the biomechanical properties of the cornea, such as the thickness, the radius of curvature, as well as the amount of the fluorescein used.

Objectives: The aim of the study was to compare IOP values measured by GAT with those measured by applanation resonance tonometry (ART).

Material And Methods: A total of 47 patients (94 eyes), including 28 patients with primary open-angle glaucoma (POAG) and 19 subjects from the control group, were examined at the Glaucoma Outpatient Clinic of the Department and Clinic of Ophthalmology at Wroclaw Medical University (Poland). The measurements of IOP were performed using GAT and a handheld version of ART. Also, the central corneal thickness (CCT) of all patients was measured.

Results: The study showed that the IOP values measured by both tonometers were comparable, but ARTacquired values were higher than GAT-obtained values both in the glaucomatous group and in the control group. CCT had little impact on mean IOP difference between GATand ART-obtained values.

Conclusions: Applanation resonance tonometry is a precise method of IOP measurement and is less affected by biomechanical properties of the cornea than GAT. Our results show that ART is a new, promising, comfortable for both patients and doctors method of IOP measurement, which, in the future, can replace GAT.

Download full-text PDF

Source
http://dx.doi.org/10.17219/acem/68559DOI Listing

Publication Analysis

Top Keywords

applanation resonance
12
iop measurement
12
intraocular pressure
8
goldmann applanation
8
handheld version
8
biomechanical properties
8
properties cornea
8
iop values
8
values measured
8
resonance tonometry
8

Similar Publications

Background: Carotid-femoral pulse wave velocity (cfPWV) is the gold standard for noninvasive arterial stiffness assessment, an independent predictor of cardiovascular disease, and a potential parameter to guide therapy. However, cfPWV is not routinely measured in clinical practice due to the unavailability of a low-cost, operator-friendly, and independent device. The current study validated a novel laser Doppler vibrometry (LDV)-based measurement of cfPWV against the reference technique.

View Article and Find Full Text PDF

Regional and Global Aortic Pulse Wave Velocity in Patients with Abdominal Aortic Aneurysm.

Eur J Vasc Endovasc Surg

March 2024

Cardiovascular Sciences, Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden; Centre for Medical Image Science and Visualisation (CMIV), Linköping University, Linköping, Sweden.

Objective: Abdominal aortic aneurysm (AAA) is commonly defined as localised aortic dilatation with a diameter > 30 mm. The pathophysiology of AAA includes chronic inflammation and enzymatic degradation of elastin, possibly increasing aortic wall stiffness and pulse wave velocity (PWV). Whether aortic stiffness is more prominent in the abdominal aorta at the aneurysm site is not elucidated.

View Article and Find Full Text PDF

Corneal Morphological and Biomechanical Changes in Thyroid-Associated Ophthalmopathy.

Cornea

May 2024

State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangzhou, China.

Purpose: This study aimed to evaluate corneal morphological and biomechanical changes in patients with thyroid-associated ophthalmopathy (TAO) and their correlations with activity and severity.

Methods: Patients diagnosed with TAO were recruited and divided into groups by activity and severity. All subjects underwent a complete ophthalmic examination, including magnetic resonance imaging.

View Article and Find Full Text PDF

The past decade has seen considerable growth in therapeutics and device technologies to treat patients with hypertension and other cardiovascular disease states. Uncoupling ventriculo-arterial interactions in these patients, however, is often complex and not adequately accounted for by arterial pressure or vascular resistance measurement alone. In reality, the global vascular load presented to the left ventricle (LV) includes both steady-state and pulsatile components.

View Article and Find Full Text PDF

Background: Breast implants are the most commonly used medical devices in plastic surgery, and capsular contracture (CC) is one of the most common complications. However, our assessment of CC is based largely on Baker grade, which is problematically subjective and affords only four possible values.

Methods: The authors performed a systematic review concluding in September of 2021 in compliance with Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!