Purpose: Wedge phantoms coupled with a CCD camera are suggested as a simple means to improve the efficiency of quality assurance for pencil beam scanning (PBS) proton therapy, in particular to verify energy/range consistency on a daily basis. The method is based on the analysis of an integral image created by a pencil beam (PB) pattern delivered through a wedge. We have investigated the reproducibility of this method and its dependence on setup and positional beam errors for a commercially available phantom (Sphinx, IBA Dosimetry) and CCD camera (Lynx, IBA Dosimetry) system.

Material And Methods: The phantom includes 4 wedges of different thickness, allowing verification of the range for 4 energies within one integral image. Each wedge was irradiated with a line pattern of clinical energies (120, 150, 180 and 230MeV). The equipment was aligned to the isocenter using lasers, and the delivery was repeated for 5 consecutive days, 4 times each day. Range was computed using the myQA software (IBA Dosimetry) and inter- and intra-setup uncertainty were calculated. Dependence of range on energy was investigated delivering the same pencil beam pattern but with energy variations in steps of ±0.2MeV for all the nominal energies, up to ±1.0MeV. Possible range uncertainties, caused by setup and positional errors, were then simulated including inclination of the phantom, pencil beam and couch shifts.

Results: Intra position setup (based on in-room laser system) shows a maximum in plane difference within 1.5mm. Range reproducibility (standard deviation) was less than 0.14mm. Setup and beam errors did not affect significantly the results, except for a vertical shift of 10mm which leads to an error in the range computation.

Conclusion: Taking into account different day-to-day setup and beam errors, day-to-day determination of range has been shown to be reproducible using the proposed system.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.zemedi.2018.02.001DOI Listing

Publication Analysis

Top Keywords

pencil beam
16
beam errors
12
iba dosimetry
12
range
8
quality assurance
8
ccd camera
8
integral image
8
beam pattern
8
setup positional
8
setup beam
8

Similar Publications

HollandPTC is an independent outpatient center for proton therapy, scientific research, and education. Patients with different types of cancer are treated with Intensity Modulated Proton Therapy (IMPT). Additionally, the HollandPTC R&D consortium conducts scientific research into the added value and improvements of proton therapy.

View Article and Find Full Text PDF

Objectives: Radiotherapy manages pancreatic cancer in various settings; however, the proximity of gastrointestinal (GI) luminal organs-at-risk (OAR) poses challenges to conventional radiotherapy. Proton beam therapy (PBT) may reduce toxicities compared to photon therapy. This consensus statement summarizes PBT's safe and optimal delivery for pancreatic tumors.

View Article and Find Full Text PDF

Objectives: The purpose of this study was to investigate the fundamental properties of spot-scanning proton beams and compare them to Monte Carlo (MC) simulations, both with and without CT calibration, using spatially diverse combinations of materials.

Methods: A heterogeneous phantom was created by spatially distributing titanium, wax, and thermocol to generate six scenarios of heterogeneous combinations. Proton pencil beams ranging in energy from 100 to 226.

View Article and Find Full Text PDF

An AI dose-influence matrix engine for robust pencil beam scanning protons therapy.

Med Phys

December 2024

National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, People's Republic of China.

Background: Rapid planning is of tremendous value in proton pencil beam scanning (PBS) therapy in overcoming range uncertainty. However, the dose calculation of the dose influence matrix (D) in robust PBS plan optimization is time-consuming and requires substantial acceleration to enhance efficiency.

Purpose: To accelerate the D calculations in PBS therapy, we developed an AI-D engine integrated into our in-house treatment planning system (TPS).

View Article and Find Full Text PDF

Purpose: A novel proton beam delivery method known as DynamicARC spot scanning has been introduced. The current study aims to determine whether the partial proton arc technique, in conjunction with DynamicARC pencil beam scanning (PBS), can meet clinical acceptance criteria for bilateral head and neck cancer (HNC) and provide an alternative to full proton arc and traditional intensity-modulated proton therapy (IMPT).

Method: The study retrospectively included anonymized CT datasets from ten patients with bilateral HNC, all of whom had previously received photon treatment.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!