Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Real time radiation dose monitoring in the cath lab may provide immediate feedback for potential dose reduction during PCI. Radiation dose monitoring to predict potential tissue injury utilizes equipment measured air Kerma at the interventional reference point (IRP) with then calculated specific tissue peak skin dose. The role of the cath lab Quality Committee is not only to assess individual high dose radiation cases but also to create processes and assess new technologies to assure radiation dose is best utilized in all cases.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/ccd.27563 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!