Objectives: The objective of this study was to assess the 12-month clinical outcomes in patients with drug-eluting stent in-stent restenosis (DES-ISR) who were either pre-dilated with non-compliant balloons (NCBA) and with additional scoring balloons (NCBA + SBA) prior to drug coated balloon (DCB) angioplasty.

Methods: This monocentric, retrospective study included patients with DES-ISR who were routinely treated over a 2-year time span. Patients with stable angina and documented ischemia or selected forms of unstable angina due to a culprit DES-ISR lesion were analyzed. The primary endpoint was the clinically driven target-lesion revascularization (TLR) rate at 12 months. Secondary endpoints included post-interventional lumen gain and late lumen loss (LLL) at 6 months.

Results: The 12-month TLR rates in 124 patients who underwent either NCBA + SBA or NCBA only group were not different (17.3%, 9/52 vs 11.6%, 8/69, P = 0.371) and low as compared to other comparable studies. The use of SBA led to equally high post minimal lumen diameters (MLD) in both treatment arms (NCBA 2.21 ± 0.33 vs NCBA + SBA 2.18 ± 0.41, P = 0.868). We did not find a significant difference in late lumen loss (LLL) between both groups (0.50 ± 0.62 mm vs 0.40 ± 0.46 mm, P = 0.468).

Conclusions: Scoring Balloon Angioplasty can safely and effectively prepare DES-ISR lesions to render them suitable for DCB angioplasty with acceptable TLR and MACE rates.

Download full-text PDF

Source
http://dx.doi.org/10.1111/joic.12506DOI Listing

Publication Analysis

Top Keywords

drug coated
8
coated balloon
8
in-stent restenosis
8
late lumen
8
lumen loss
8
loss lll
8
score balloons
4
balloons enhance
4
enhance outcomes
4
outcomes drug
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!