Failure of a "foolproof" pin-index medical pipeline system.

JA Clin Rep

Surgical Center, Kumamoto University Hospital, 1-1-1 Honjo, Chuo-ku, Kumamoto, 860-8556 Japan.

Published: August 2016

Background: The pin-index medical gas pipeline system, which complies with Japan Industrial Standard (JIS), is considered to be "foolproof" and is widely used in Japan to avoid medical gas misconnections.

Case Presentation: The wall-mounted gas outlet used in our hospital (NSV outlet, CENTRAL UNI, Co., Ltd., Tokyo, Japan) contains multiple sockets, which connect to hoses with gas-specific pins. Each socket is covered with a gas-specific plastic pin guide, which is considered to make the system foolproof; i.e., to prevent misconnections. However, while checking an anesthesia machine in accordance with the guidelines developed by the Japanese Society of Anesthesiologists, an anesthesiologist found that one of the gas-specific plastic pin guides covering the wall-mounted gas outlets had disappeared; and hence, the gas outlet system was no longer foolproof. A subsequent verification test performed by engineers of the system's manufacturer revealed that the plastic pin guides could be dislodged by applying 29.4 N of vertical force.

Conclusions: It is important to check that gas outlet systems are functioning in a gas-specific manner when they are used for clinical purposes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5818779PMC
http://dx.doi.org/10.1186/s40981-016-0044-7DOI Listing

Publication Analysis

Top Keywords

gas outlet
12
plastic pin
12
pin-index medical
8
pipeline system
8
medical gas
8
wall-mounted gas
8
gas-specific plastic
8
pin guides
8
gas
6
failure "foolproof"
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!