Energy-dependent dead-time correction in digital pulse processors applied to silicon drift detector's X-ray spectra.

J Synchrotron Radiat

Instituto de Física, Universidade de São Paulo, Rua do Matão 1371, Cidade Universitária, CEP 05508-090, São Paulo, SP, Brazil.

Published: March 2018

AI Article Synopsis

Article Abstract

Dead-time effects in X-ray spectra taken with a digital pulse processor and a silicon drift detector were investigated when the number of events at the low-energy end of the spectrum was more than half of the total, at counting rates up to 56 kHz. It was found that dead-time losses in the spectra are energy dependent and an analytical correction for this effect, which takes into account pulse pile-up, is proposed. This and the usual models have been applied to experimental measurements, evaluating the dead-time fraction either from the calculations or using the value given by the detector acquisition system. The energy-dependent dead-time model proposed fits accurately the experimental energy spectra in the range of counting rates explored in this work. A selection chart of the simplest mathematical model able to correct the pulse-height distribution according to counting rate and energy spectrum characteristics is included.

Download full-text PDF

Source
http://dx.doi.org/10.1107/S1600577517018318DOI Listing

Publication Analysis

Top Keywords

energy-dependent dead-time
8
digital pulse
8
silicon drift
8
x-ray spectra
8
counting rates
8
dead-time correction
4
correction digital
4
pulse processors
4
processors applied
4
applied silicon
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!