Purpose: To quantify the patient-specific imaging dose for real-time tumour monitoring in the lung during respiratory-gated stereotactic body radiotherapy (SBRT) in clinical cases using SyncTraX.

Methods And Materials: Ten patients who underwent respiratory-gated SBRT with SyncTraX were enrolled in this study. The imaging procedure for real-time tumour monitoring using SyncTraX was simulated using Monte Carlo. We evaluated the dosimetric effect of a real-time tumour monitoring in a critical organ at risk (OAR) and the planning target volume (PTV) over the course of treatment. The relationship between skin dose and gating efficiency was also investigated.

Results: For all patients, the mean D50 to the PTV, ipsilateral lung, liver, heart, spinal cord and skin was 118.3 (21.5-175.9), 31.9 (9.5-75.4), 15.4 (1.1-31.6), 10.1 (1.3-18.1), 25.0 (1.6-101.8), and 3.6 (0.9-7.1) mGy, respectively. The mean D2 was 352.0 (26.5-935.8), 146.4 (27.3-226.7), 90.7 (3.6-255.0), 42.2 (4.8-82.7), 88.0 (15.4-248.5), and 273.5 (98.3-611.6) mGy, respectively. The D2 of the skin dose was found to increase as the gating efficiency decreased.

Conclusions: The additional dose to the PTV was at most 1.9% of the prescribed dose over the course of treatment for real-time tumour monitoring. For OARs, we could confirm the high dose region, which may not be susceptible to radiation toxicity. However, to reduce the skin dose from SyncTraX, it is necessary to increase the gating efficiency.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aab242DOI Listing

Publication Analysis

Top Keywords

real-time tumour
20
tumour monitoring
20
skin dose
12
gating efficiency
12
patient-specific imaging
8
dose
8
imaging dose
8
dose real-time
8
monitoring lung
8
course treatment
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!