AI Article Synopsis

  • VOCs are important indicators in various fields, including climate change, indoor air quality, food safety, and health, but current measurement methods are complex and costly due to the need for standard solutions.
  • The paper introduces portable generators for producing trace-level VOC gas standards, which simplify calibration of VOC detectors and eliminate the need for pre-prepared calibration standards in cylinders.
  • Two devices developed by national metrology institutes (INRIM and METAS) use different technologies to generate reference gas mixtures in a precise and efficient manner for practical lab applications.

Article Abstract

Volatile organic compounds (VOCs) in gas mixtures at trace level (nmol/mol) are routinely measured by chemical and biochemical laboratories as climate indicators, indoor air quality pollutants from building materials emissions, contaminants in food and beverages, and biomarkers in body fluids (blood, urine, breath) of occupational exposure or human diseases. Current analytical instruments used for measurements are gas chromatographs equipped with various injector and detector configurations. The assurance of measurement quality is done by using a huge amount of certified liquid VOC standard solutions (or gaseous VOC standard cylinders) with multiple dilutions to reach the required trace level. This causes high standard uncertainty in instrument calibrations, high cost, and high consumption of analysis and laboratory personal time. In this paper, we present the implementation of portable generators producing VOC gas standards at trace level for automatic and direct calibration of VOC detectors employed in various contexts, removing the need for preparation of matrix calibration standards in cylinders. Two compact devices in-house developed by two national metrology institutes-the Istituto Nazionale di Ricerca Metrologica (INRIM) and the Federal Institute of Metrology (METAS)-are here used to dynamically generate reference gas mixtures in an SI traceable way. The two devices are based on different technologies: diffusion and permeation, for INRIM and METAS, respectively. A metrological characterization is given and the practical implementation at chemical and biochemical laboratories is discussed. Graphical abstract Onsite calibration with transportable generation system with similar performances to primary laboratory devices.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00216-018-0935-8DOI Listing

Publication Analysis

Top Keywords

chemical biochemical
12
biochemical laboratories
12
trace level
12
compact devices
8
gas mixtures
8
voc standard
8
voc
5
devices generation
4
generation reference
4
trace
4