Background: There is no evidence from randomized trials for the benefit of routine non-compliant balloon (NCB) post-dilation after stent deployment. Despite being the gold standard, intravascular ultrasound is infrequently performed due to time and cost constraints and a suitable alternative technology is required for routine assessment of stent expansion. The purpose of this study was to assess the contribution of NCB post-dilation in optimizing contemporary stents by using digital stent enhancement (DSE).
Methods: We treated 120 patients with stent insertion and assessed the stents with DSE before and after NCB use. Optimal expansion was defined as the minimum stent diameter (MSD) ≥ 90% of the nominal stent diameter, an adaptation of the MUSIC and POSTIT trial criteria. Stent deployment was performed at 12 atm pressure followed by routine NCB post-dilation at ≥ 14 atm.
Results: The mean reference diameter on QCA was 2.75 mm (SD 0.63) and mean stent diameter was 3.15 mm (SD 0.46). At a mean stent deployment pressure of 11.7 atm (SD 2.4), only 21% of stents were optimally expanded. After NCB inflation at a mean of 16.9 atm (SD 2.8), MSD increased by 0.26 mm (SD 0.24), optimal stent expansion increased from 21% to 58% and mean stent symmetry ratio increased from 0.83 to 0.87 (p < 0.0001).
Conclusions: Contemporary stents are sub-optimally expanded in the majority of cases after standard deployment compared with nominal sizes. Adjunctive NCB post-dilation optimized an additional 37% of stents. DSE analysis can assist in qualitative and quantitative stent assessments and can potentially facilitate a selective NCB post-dilation strategy to achieve optimal stent expansion.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5801274 | PMC |
http://dx.doi.org/10.1016/j.ijchv.2014.03.006 | DOI Listing |
Int J Cardiol Heart Vessel
June 2014
Department of Cardiology, The Canberra Hospital, Australia.
Background: There is no evidence from randomized trials for the benefit of routine non-compliant balloon (NCB) post-dilation after stent deployment. Despite being the gold standard, intravascular ultrasound is infrequently performed due to time and cost constraints and a suitable alternative technology is required for routine assessment of stent expansion. The purpose of this study was to assess the contribution of NCB post-dilation in optimizing contemporary stents by using digital stent enhancement (DSE).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!