Intraoperative Endotracheal Cuff Pressure Study: How Education and Availability of Manometers Help Guide Safer Pressures.

Mil Med

Department of Anesthesiology, San Antonio Uniformed Services Health Education Consortium, 3551 Roger Brooke Dr., JBSA Fort Sam Houston, TX.

Published: September 2018

Introduction: Endotracheal intubation is a medical procedure that is often indicated in both the perioperative and critical care environments. Cuffed endotracheal tubes (ETT) allow for safer and more efficient delivery of positive pressure ventilation, as well as create a barrier to reduce the risk of micro-aspiration and anesthetic pollution in the operating room environment. Over-inflation of the endotracheal cuff can lead to serious and harmful sequelae. This study aimed to assess if departmental education paired with ready access to a manometer to assess cuff pressure would result in an improvement in the proportion of ETT cuff pressures in the goal range.

Materials And Methods: A quality improvement study was conducted at the San Antonio Military Medical Center (SAMMC; Department of Defense hospital in San Antonio, TX). The initiative was divided into three key periods: pre-implementation, implementation, and post-implementation. During the pre-implementation period, ETT cuff pressures were obtained to assess the baseline state of ETT cuff pressures for patients in the operating room; the proportion of in-range (20-30 cmH2O) pressures was calculated. During the implementation phase, operating rooms were equipped with manometers and anesthesia departmental education was performed regarding the use of the manometers. Three months later, post-implementation cuff pressures were measured in the OR, and the proportion of in-range pressures was again calculated.

Results: The pre-implementation data showed an average cuff pressure of 48.92 cmH2O and a median of 38.5 cmH2O. Of the 100 pre-implementation pressures, 20 were in the goal range. Post-implementation data had an average cuff pressure of 41.96 cmH2O and a median of 30 cmH2O. A chi-squared test of pressures in the safe range from the pre-implementation versus post-implementation values yielded a highly significant p-value of 0.0003.

Conclusion: The data from this study clearly demonstrated a statistically significant improvement in the proportion of in-range cuff pressures following the quality improvement initiative. This study supports the use of department-wide education and the availability of manometers in each OR to yield safer cuff pressures for intubated patients. This study did not aim to determine an optimal ETT cuff pressure, but utilized data already available to determine a safe cuff pressure. Further research needs to be performed to assess whether routine monitoring of cuff pressure results in improved patient outcomes.

Download full-text PDF

Source
http://dx.doi.org/10.1093/milmed/usx127DOI Listing

Publication Analysis

Top Keywords

cuff pressure
28
cuff pressures
24
ett cuff
16
cuff
14
proportion in-range
12
pressures
11
endotracheal cuff
8
pressure
8
education availability
8
availability manometers
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!