High-performance liquid chromatography (HPLC) was used to separate oxamyl from other pesticides in drinking water and tomato paste. The eluate emerging from the column tail was mixed with an alkaline solution of Co in EDTA and irradiated with UV light to induce photolysis of the carbamate in order to obtain free radicals and other reactive species that oxidize luminol and produce chemiluminescence (CL) as a result. The intensity of the CL signal was monitored in the form of chromatographic peaks. Under the optimum operating conditions for the HPLC-UV-CL system, the analyte concentration was linearly related to peak area. The limit of detection as determined in accordance with the IUPAC criterion was 0.17 mg L. Oxamyl was successfully extracted with recoveries of 88.7-103.1% from spiked tomato paste by using a simple QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) sample preparation approach. Similar recoveries were obtained from drinking water samples spiked with oxamyl concentrations above the LOD. The proposed method is a simple, fast, accurate choice for quantifying this pesticide.