Background: Unfavorable anatomy is one of the major limitations of endovascular aortic aneurysm repair (EVAR) with specifically adverse proximal neck morphology excluding many patients from receiving the standard endograft devices. Thoracic tube endografts have been used to overcome the issue of wide infrarenal necks, either as a sole device (single tubes or double tubes using the trombone technique) or as the proximal part of a bifurcated device fixed to the aortic bifurcation or infrarenally oriented. However, custom-made large proximal diameter bifurcated endograft designs have never been used.

Methods: We present the case of a 56-year-old man with Marfan syndrome, suffering abdominal aortic aneurysm (AAA) enlargement from a type Ib endoleak after previous EVAR with 2 Endofit tube endografts (trombone technique). He was considered unfit for open surgery while possible alternatives such as fenestrated endovascular AAA repair and chimney technique were excluded.

Results: The patient was treated with a custom-made 44-mm proximal diameter, bifurcated Bolton Medical Treovance device with technical and clinical success. No immediate or perioperative complications were noted. Follow-up after 6 months showed graft patency and no endoleak of any type.

Conclusions: This alternative technique for hostile proximal neck management is promising and needs long-term follow-up; an issue which is discussed within the broader context of custom-made device regulations.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.avsg.2017.10.029DOI Listing

Publication Analysis

Top Keywords

aortic aneurysm
12
trombone technique
12
abdominal aortic
8
proximal neck
8
tube endografts
8
proximal diameter
8
diameter bifurcated
8
technique
5
proximal
5
custom-made
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!