MCNP5 evaluation of photoneutron production from the Alexandria University 15 MV Elekta Precise medical LINAC.

Appl Radiat Isot

Physics and Chemistry Department, Faculty of Education, Alexandria University, Alexandria, Egypt. Electronic address:

Published: May 2018

AI Article Synopsis

Article Abstract

Photoneutron production, and the dose equivalent, in the head assembly of the 15 MV Elekta Precise medical linac; operating in the faculty of Medicine at Alexandria University were estimated with the MCNP5 code. Photoneutron spectra were calculated in air and inside a water phantom to different depths as a function of the radiation field sizes. The maximum neutron fluence is 3.346×10 n/cm-e for a 30×30 cm field size to 2-4 cm-depth in the phantom. The dose equivalent due to fast neutron increases as the field size increases, being a maximum of 0.912 ± 0.05 mSv/Gy at depth between 2 and 4 cm in the water phantom for 40×40 cm field size. Photoneutron fluence and dose equivalent are larger to 100 cm from the isocenter than to 35 cm from the treatment room wall.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2018.01.036DOI Listing

Publication Analysis

Top Keywords

dose equivalent
12
field size
12
photoneutron production
8
alexandria university
8
15 mv elekta
8
elekta precise
8
precise medical
8
medical linac
8
water phantom
8
mcnp5 evaluation
4