In this paper, we present a method that uses a combination of experimental and modeled data to assess properties of x-ray beam measured using a small-animal spectral scanner. The spatial properties of the beam profile are characterized by beam profile shape, the angular offset along the rotational axis, and the photon count difference between experimental and modeled data at the central beam axis. Temporal stability of the beam profile is assessed by measuring intra- and interscan count variations. The beam profile assessment method was evaluated on several spectral CT scanners equipped with Medipix3RX-based detectors. On a well-calibrated spectral CT scanner, we measured an integral count error of 0.5%, intrascan count variation of 0.1%, and an interscan count variation of less than 1%. The angular offset of the beam center ranged from 0.8° to 1.6° for the studied spectral CT scanners. We also demonstrate the capability of this method to identify poor performance of the system through analyzing the deviation of the experimental beam profile from the model. This technique can, therefore, aid in monitoring the system performance to obtain a robust spectral CT; providing the reliable quantitative images. Furthermore, the accurate offset parameters of a spectral scanner provided by this method allow us to incorporate a more realistic form of the photon distribution in the polychromatic-based image reconstruction models. Both improvements of the reliability of the system and accuracy of the volume reconstruction result in a better discrimination and quantification of the imaged materials.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5849855 | PMC |
http://dx.doi.org/10.1002/acm2.12260 | DOI Listing |
Sensors (Basel)
January 2025
Department of Civil Engineering, Myongji College, Seoul 03656, Republic of Korea.
Conventional approaches for the structural health monitoring of infrastructures often rely on physical sensors or targets attached to structural members, which require considerable preparation, maintenance, and operational effort, including continuous on-site adjustments. This paper presents an image-driven hybrid structural analysis technique that combines digital image processing (DIP) and regression analysis with a continuum point cloud method (CPCM) built on a particle-based strong formulation. Polynomial regressions capture the boundary shape change due to the structural loading and precisely identify the edge and corner coordinates of the deformed structure.
View Article and Find Full Text PDFBrachytherapy
January 2025
BC Cancer Kelowna, Kelowna, British Columbia.
Purpose: High dose rate (HDR) brachytherapy is increasingly adopted for dose escalation in prostate cancer treatment. We report the clinical efficacy and toxicity of HDR prostate brachytherapy combined with external beam radiotherapy (EBRT) and evaluate the predictability of the biochemical definition of cure of 4-year PSA ≤0.2 ng/mL for failure free survival (FFS).
View Article and Find Full Text PDFPhys Med Biol
January 2025
Department of Radiation Oncology, The University of Kansas Medical Center, Department of Radiation Oncology, University of Kansas Medical Center, USA, Kansas City, Kansas, 66160-8500, UNITED STATES.
Proton minibeam radiation therapy (pMBRT) dose profile is characterized by highly heterogeneous dose in the plane perpendicular to the beam direction and rapidly changing depth dose profiles. Typically, dose measurements are benchmarked against in-house Monte Carlo simulation tools. It is essential to have a treatment planning system (TPS) that can accurately predict pMBRT doses in tissue and be available via commercial platform for preclinical and clinical use.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Radiation Oncology, Duke University, North Carolina, USA.
Background: The electronic compensation (ECOMP) technique for breast radiation therapy provides excellent dose conformity and homogeneity. However, the manual fluence painting process presents a challenge for efficient clinical operation.
Purpose: To facilitate the clinical treatment planning automation of breast radiation therapy, we utilized reinforcement learning (RL) to develop an auto-planning tool that iteratively edits the fluence maps under the guidance of clinically relevant objectives.
Sci Rep
January 2025
Department of Oral and Maxillofacial Surgery, Faculty of Medicine, Medical Center - University of Freiburg, University of Freiburg, Freiburg, Germany.
This experimental phantom study investigates current standard of care protocols in cone beam computed tomography (CBCT), energy-integrating-detector (EID) CT, and photon-counting-detector (PCD) CT regarding their potential in delineation of dental root canals. Artificial accessory canals (diameters: 1000, 600, 400, 300 and 200 μm) were drilled into three bovine teeth mounted on a bovine rib as a jaw substitute. The phantom was scanned in two dental CBCTs, two EID-CTs and a PCD-CT using standard clinical protocols.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!