Addition of internal electrodes is beneficial for focused bioimpedance measurements in the lung.

Physiol Meas

Philips Chair for Medical Information Technology, Helmholtz Institute for Biomedical Engineering, RWTH Aachen University, Aachen, Germany.

Published: March 2018

Objective: Bioimpedance measurements such as bioimpedance spectroscopy (BIS) or electrical impedance tomography (EIT) are used in many biomedical applications. While BIS measures and analyzes the impedance in a frequency range at constant electrode positions, EIT aims to reconstruct images of the conductivity distribution from multiple measurements at different electrode positions. Our aim is to add spatial information to tetrapolar BIS measurements by using electrode positions that focus measurements on desired regions of interest. In this paper, we aim to investigate, whether internal electrodes that can be integrated into breathing or gastroesophageal tubes, can improve the local sensitivity of bioimpedance spectroscopy measurements.

Approach: We present the results of a simulation study, in which we investigated more than 4 M different electrode configurations on their ability to monitor specific regions of interest (ROI) in the lung. Based on the sensitivity, which describes the impact of a conductivity change on the measured impedance, we define three main criteria which we use to evaluate our simulation results: the selectivity [Formula: see text], which describes the impact of a conductivity change inside the region of interest compared to a conductivity change outside the ROI; the homogeneity [Formula: see text], which describes the distribution of the sensitivity inside the ROI; and the absolute impedance contribution ratio [Formula: see text], which describes the contribution of the ROI to the measured impedance.

Main Results: Depending on the region of interest, electrode configurations using internal electrodes are between 9.8 % and 90 % better with respect to these criteria than configurations using external electrodes only.

Significance: The combination of internal and external electrodes improves the focusing ability of tetrapolar impedance measurements on specific lung regions, which may be especially beneficial for lung monitoring in intensive care.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6579/aaad45DOI Listing

Publication Analysis

Top Keywords

internal electrodes
12
electrode positions
12
conductivity change
12
[formula text]
12
text] describes
12
bioimpedance measurements
8
bioimpedance spectroscopy
8
measurements electrode
8
regions interest
8
electrode configurations
8

Similar Publications

Enabling ultra-flexible inorganic thin-film-based thermoelectric devices by introducing nanoscale titanium layers.

Nat Commun

January 2025

School of Chemistry and Physics, ARC Research Hub in Zero-emission Power Generation for Carbon Neutrality, and Centre for Materials Science, Queensland University of Technology, Brisbane, QLD, Australia.

Here, we design exotic interfaces within a flexible thermoelectric device, incorporating a polyimide substrate, Ti contact layer, Cu electrode, Ti barrier layer, and thermoelectric thin film. The device features 162 pairs of thin-film legs with high room-temperature performance, using p-BiSbTe and n-BiTeSe, with figure-of-merit values of 1.39 and 1.

View Article and Find Full Text PDF

Objective: Electrical Impedance Tomography (EIT) is a non-invasive technique used for lung imaging. A significant challenge in EIT is reconstructing images of deeper thoracic regions due to the low sensitivity of boundary voltages to internal conductivity variations. The current injection pattern is decisive as it influences the current path, boundary voltages, and their sensitivity to tissue changes.

View Article and Find Full Text PDF

Context: The two-dimensional graphene/MoTe heterostructure holds extensive potential applications in optoelectronic devices, sensors, and catalysts. To expand its optical applications, this study systematically investigates the adsorption stability of metal atoms (Au, Pt, Pd, and Fe) on the graphene/MoTe and their influence on its optoelectronic properties employing first-principles methods. The findings indicate that after the adsorption of Au and Pd, the structure retains its direct bandgap properties, while the adsorption of Pt and Fe exhibits indirect bandgap characteristics.

View Article and Find Full Text PDF

Soft and stretchable strain sensors are crucial for applications in human-machine interfaces, flexible robotics, and electronic skin. Among these, capacitive strain sensors are widely used and studied; however, they face challenges due to material and structural constraints, such as low baseline capacitance and susceptibility to external interference, which result in low signal-to-noise ratios and poor stability. To address these issues, we propose a U-shaped electrode flexible strain sensor based on liquid metal elastomer (LME).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!